(7) When Treatment is Finished.

· Release the Foot Control and attach the handpiece to the 

handpiece holder. (Fig. 16)

· Turn off the Power Switch.
· Close the dental unit's water valve.
· Remove the Tip after use. (Refer. 4. (2)  Tip Connection)

When the Control Unit is shut down, the last mode settings in use are automatically retained in memory. When you 
power on the system for the next procedure, the system will reactivate in the same configuration as when shut 
down.

NOTICE

Fig. 16

Fig. 17

*

Memory Function

The Control Unit can be preset for Water Mode, power 
setting and operating mode. To set the desired preset 
conditions into memory follow the procedure below.
1) Set the desired levels.
2) Push the Memory Key for about 2 seconds or more, the 

memory light will illuminate and the control unit will 
produce a beep. The memory is set. (Fig. 17)

*

Safety Function

· When the operating mode is set to G and the power setting is set to Max the Control Unit will automatically reduce 

power if operated at these settings for more than 10 minutes. (The output display moves 3 lamps from "Max" to "4".)

To reset the safety function....

Before a lapse of 10 minutes or when the power weakens after a lapse of 10 minutes, release the Foot Control. The 
above safety function is reset. Depressing the Foot Control again repeats the same operation.

· If the Power Switch is turned on while the Foot Control is depressed accidentally ( by an object, etc.), beeping starts to 

call attention.

To reset the alarm....

Release the Foot Control (remove the object, etc.).

To enter new settings....

After confirming the settings of Water Mode, Power Setting and Operating Mode, push the memory key for approx. 2 
seconds. (The memory light will illuminate.) The old memory is cleared and new settings are programmed.

Memory Key

Memory Lamp

When the Tip is mounted to the handpiece, install the Tip Cover 
S for safety.
· Hold the Tip Cover S as shown and insert it all the way to the 

Tip-Handpiece joint.

· To remove, hold the Tip Cover S and the handpiece as shown 

and pull straight out.

5. How to Use Tip Cover S

Fig. 18

Handpiece

Slit

WARNING

Carefully insert the tip into the Tip Cover S. Avoid injuring the fingers.

Do not use the Tip Cover S as a Tip change tool.

CAUTION

· Use the tip holder for tips removed from the handpiece.
· The tip holder is autoclavable and can hold up to 5 types at 

once. To autoclave, tilt the tips in the direction of the arrow in 
fig.19.

6. Tip Holder

· Autoclave sterilization is recommended.
· Sterilization is required first time you use and after each patient as noted below.
· The Tip, Handpiece, Tip Wrench, Tip Cover S, and Tip Holder can be autoclaved.

7. Sterilization

Do not autoclave any parts (the control unit, power cord, foot control, water tube, irrigation connector, handpiece 
cord including the cover, lamp, O-ring, tip card, and spanner wrench (5 X 8) ) other than those that can be 
subjected to autoclave sterilization.

CAUTION

Fig. 19

Slit

Tip

Tip Cover S

Tip-handpiece Joint

Handpiece

Handpiece Holder

Handpiece Holder

Tip

Tip Holder

Summary of Contents for Varios 560

Page 1: ...read this Operation Manual carefully before use and file for future reference Multi Function Ultrasonic Scaler OM E0248E 004 10 03 01 S The EU directive 93 42 EEC was applied in the design and production of this medical device ...

Page 2: ...evice Safety instructions are classified as follows in accordance with the seriousness of the risk Use by medical professional such as doctor or dental hygienist is intended Do not unplug the Power Cord with wet hands to avoid electric shock Be sure to prevent water on the Control Unit because it may result in short circuit and electric shock Do not allow any impacton the Control Unit Do not drop T...

Page 3: ...on and overheating outside the patient s oral cavity before use If any abnormalities are found stop using immediately and contact your dealer Should this products function abnormally cease operation immediately and returen it to the dealer for repair Do not force or pull on Power Cord and or Handpiece Cord It could cause disconnection Be sure to attach NSK genuine tips when using NSK Varios ultras...

Page 4: ... Control Plug Water Filter Case Water Filter Case Water Tube Water Tube Water Connector Water Connector 4 Handpiece Cord Connection Align the marks on the Control Unit and the Handpiece Cord Plug Gently push the Plug into the Connector CAUTION 1 Water System Setup Carefully check all Water Supply Connections prior to starting procedure Open the dental unit s water valve Set water pressure between ...

Page 5: ...urface can be injured 2 Set the Power Level with the Power Level Key on the Front Panel The lamps in the Power Display Panel will indicate the selected power level Check to make sure the Power Level is set in the appropriate range for the selected Tip Fig 12 NOTICE Continuous pushing of the Power Level Key will cause the Power Level Display to climb When only water is desired without oscillating t...

Page 6: ...lly by an object etc beeping starts to call attention To reset the alarm Release the Foot Control remove the object etc To enter new settings After confirming the settings of Water Mode Power Setting and Operating Mode push the memory key for approx 2 seconds The memory light will illuminate The old memory is cleared and new settings are programmed Memory Key Memory Lamp When the Tip is mounted to ...

Page 7: ...p replacement is not necessary Yellow Wear of 1mm Tip is showing some wear Tip replacement is recommended Red Wear of 2mm Tip is badly worn Tip replacement is necessary Fig 20 OK NO Tip Card Handpiece Neck The Tip Card can be used to check the following tips G1 G4 G5 G6 G8 P1 P1D P10 and P20 Efficiency 1mm 2mm 25 50 Decrease Decrease 2 Changing Water Filter 1 Close the water valve of the dental uni...

Page 8: ...oting No Poor vibration The tip is bent or broken The tip walks out Noise from the handpiece The handpiece is overheating The Power Lamp does not light even if the Power Switch is ON The tip does not generate vibration in spite of depressing the Foot Control The Power Cord or the Jack is disconnected The internal Fuse has burned out The tip is not tightened firmly Worn Tip Output has not been corre...

Page 9: ... turns on and off Tip oscillates but light does not turn on Lamp is correctly and securely mounted in the socket but light does not turn on Power output is set at maximum at G mode Continuous operation for a long time at high mode G General Operation in a high temperature room Beeping while power on Beeping while stopping vibration of Tips The Water Adjustment Dial is closed Disconnected irrigatio...

Page 10: ... measures may be necessary such as reorienting or relocating the Varios560 Varios560LUX b Over the frequency range 150kHz to 80MHz field strengths should be less than 3 V m Immunity test IEC60601 test level Electromagnetic environment guidance Conducted RF IEC61000 4 6 Radiated RF IEC61000 4 3 3Vrms 150 kHz to 80MHz 3V m 80MHz to 2 5 GHz Compliance level 3Vrms 3V m Portable and mobile RF communicat...