20. EMC Information
The instrument fulfils the requirements according to EN 60601-1-2 regarding emission and immunity. If sensitive electronic equipment is affected by the
instrument, try to increase the distance to such equipment. The charger should not be connected during measurements.
Guidance and manufacturer's declaration - Electromagnetic Emissions.
Osseo 100 is intended for use in the electromagnetic environment specified below.
Emissions tests
Compliance
Electromagnetic environment - guidance
RF emissions C1SPR11
Group 1
Osseo 100 uses RF energy only for its internal function.
RF emissions C1SPR11
Class B
Osseo100 Rechargeable battery operated device
Harmonic emissions IEC61000-3-2
Not applicable
Voltage fluctuations/flicker emissions IEC61000-3-3
Not applicable
Guidance and manufacturer's declaration - Electromagnetic Immunity Test Levels
Osseo100 is intended for use in the electromagnetic environment specified below.
Immunity test
EMC standard or test method
Professional healthcare facility environment
Electrostatic discharge (ESD)
IEC61000-4-2
± 8kV contact
± 2 kV ± 4 kV ± 8 kV ± 15 kV air
Radiated RF EM fields
IEC61000-4-3
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
Proximity fields form RF wireless communications equipment IEC61000-4-3
30 cm minimum separation distance from radio transmitter
Rated power frequency magnetic fields
IEC61000-4-8
30 A/m
50 Hz or 60 Hz
Electrical fast transient/burst
IEC 61000-4-4
± 2kV
100 kHz repetition frequency
Surges Line-to-line, Surges Line-to-ground
IEC 61000-4-5
± 0.5, ± 1 kV, ± 2 kV
Conducted disturbances induced by RF fields
IEC61000-4-6
3V
0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Voltage dips, Voltage interruptions and Electrical transient
condition along supply lines
IEC 61000-4-11
5% UT, 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle
And 70 % UT; 25/30 cycles (50/60Hz)
Single phase: at 0°
0 % UT; 250/300 cycle (50/60 Hz)