background image

14

15

At the bottom of the handpiece stand, tools 
necessary for handpiece maintenance and a 
spare chuck (optional) can be mounted. 
(Fig. 16)

13. Handpiece Stand

14. Specification

Fig.16

Food Pedal

Model

Weight

FC-64
460g

Handpiece Stand

Model

Weight

Handpiece Stand
120g

Control Unit

Model

Power supply

Weight

Model

Ultimate XL-K
Ultimate XL-D
AC120V 50/60 Hz 46VA
AC230V 50/60 Hz 46VA
Ultimate XL-K
Ultimate XL-D
Ultimate XL-K
Ultimate XL-D

NE231
NE232

3.2kg
3.3kg
W90 × D270× H230 mm
W81 × D270× H230 mm

Motor Handpiece

1 ) Torque Type

Model

Speed

Max. Torque

Weight

Dimensions

Code Length

UMXL-T(P) 2M
1,000~50,000 min

-1

8.7N·cm
215g (exclude the code)
L151× ø 29mm
2.0m

2 ) Compact Type

Model

Speed

Max. Torque

Weight

Dimensions

Code Length

UMXL-C 2M
1,000~50,000 min

-1

6 N·cm 
181g (exclude the code)
L144 x  ø 27mm 
2.0m 

Trouble

Cause / Check

Remedy

Please check the following points before sending back instruments for repair.

<Unit / Motor>

15. Troubleshooting

English

No error code 

indication

Error code E0 

appears.

Error code E1 

appears.

Error code E2 

appears.

The power plug is disconnected.

The fuse is blown.

Power Switch is faulty.

Power on the Control Unit

within pressing the Foot Pedal.

The connection of the foot control 

cord plug is loose.
Maintenance mode (4) “Pd”: 

Check to see if the foot control 

operates normally.

Turn on the power again.

Turn on the power again.

If you have two or more units, 

replace the motor and the motor 

cord and check the operation.

The motor cord is disconnected.

If you have two or more units, 

replace the motor and the motor 

cord and check the operation.

Insert the power plug correctly.
Replace it with a specified fuse. If the 

reason the fuse has blown is unknown, 

ask for an inspection.
Ask for repair.
Power on the Control Unit without 

pressing the Foot Pedal. When it does 

not work definitely, take off the Foot 

Pedal cord plug from the Control Unit 

and Power it on. If it does not operate 

correctly, ask for your dealer.

Even after you tried above and the 

Control Unit could not operate 

normally, contact your dealer for repair.
Connect the foot control cord plug 

correctly.
If the foot control dose not operates 

normally, ask for repair of the foot 

control or replace it with a new one.
If the same error code appears, ask for 

repair of the unit.
If it operates normally, the error display 

is temporarily due to overload, which is 

not a problem.
If it operates normally after replacing 

the motor and the motor cord, the 

motor and/or the motor cord may be 

shorted. Ask for repair of the motor 

and/or the motor cord. If the same 

error code still appears after replacing, 

ask for repair of the unit.
Connect the motor cord correctly.
If it operates normally after replacing 

the motor and the motor cord, the 

motor and/or the motor cord may be 

severed. Ask for repair of the motor 

and/or the motor cord. If the same 

error code still appears after replacing, 

ask for repair of the unit.

The motor and handpiece do not run, or Reset Lamp lights.

Pilot Lamp does not 

light

Control Unit display 

“—“and “Setting 

Rotation Speed” 

alternately.

Summary of Contents for NE231

Page 1: ...OPERATION MANUAL Micromotor for Laboratory Work English ...

Page 2: ...ine Option 6 Protective Circuit 7 Memory Function 8 Error Code 9 Replacement of Fuse 10 Maintenance Mode 11 Vacuum coupled Mode 12 Handling of Motor and Handpiece 13 Handpiece Stand 14 Specification 15 Troubleshooting 16 Disposing Product 1 5 6 7 8 8 8 9 9 10 10 11 12 14 14 15 18 Contents WARNING When using electric tools basic safety precautions should always be followed to reduce the risk of fir...

Page 3: ...zed cord will cause a drop in line voltage resulting in loss of power and overheating Table below shows the correct size to use depending on cord length and nameplate ampere rating If in doubt use the next heavier gauge The smaller the gauge number the heavier the cord 9 Install equipment so that the power supply cord can be pulled out without hindrance in event of emergency 1 For your own safety ...

Page 4: ...oto may cause damage 25 Do not turn the bur lock Chuck Control Ring while the handpiece is rotating 26 Do not rotate the motor when the bur lock Chuck Control Ring is at OPEN position or a bur is not mounted in the chuck It may cause the motor handpiece disconnection or sudden heat generation 27 Activation of Circuit Breaker means too much load is applied to the motor beyond the capacity the motor...

Page 5: ...t to the switch will light up If the desired speed is set with the Knee Controller or the Pedal and a uniform speed can be maintained for one second the motor will continue running at the desired speed if the Knee Controller or the Pedal is released To cancel it depress the Knee Controller or the Pedal again or press Auto Cruise Switch again 6 7 1 Mounting of the Unit 1 ULTIMATE XL K Bore a hole w...

Page 6: ... given to the rotation direction 7 Memory Function When the motor is stopped due to some trouble such as a failure overload wire breakage or misuse Speed Display displays the error code for checking the status of the unit and understanding the cause of the trouble easily 8 Error Code Fig 7 CAUTION Use of 40 000min 1 rpm or more is allowable only when the bur manufacturer or dealer specifies that t...

Page 7: ... To release Maintenance Mode turn Power Switch off and switch on again 1 oP Switch check operation check Press the switches on the panel and the right and or left lamps will light to check to see if the switches operate normally 2 dP Display check Press Forward Reverse Selector Switch and the lamps will light one by one to check to see if they operate normally To cancel this check press Forward Re...

Page 8: ...tc accumulate in the chuck and the bur is caught insecurely causing runout OPEN LOCK Loosen Loosen Loosen Fig 11 Loosen Loosen Loosen 3 Insertion of Chuck Thinly apply oil before insertion Open the Chuck Control Ring insert the dummy bur or the bur in use into the chuck and turn the chuck clockwise by hand until it stops Then lock the Chuck Control Ring and the chuck could hold the bur securely Fi...

Page 9: ... cord and check the operation The motor cord is disconnected If you have two or more units replace the motor and the motor cord and check the operation Insert the power plug correctly Replace it with a specified fuse If the reason the fuse has blown is unknown ask for an inspection Ask for repair Power on the Control Unit without pressing the Foot Pedal When it does not work definitely take off th...

Page 10: ...unit Ask for repair of the turbine adapter and or the Ultimate XL K D unit The motor and handpiece do not run or Reset Lamp lights Error code E3 appears Error code E4 appears Error code E5 appears Error code E6 appears Error code E8 appears The motor cord is disconnected Maintenance mode 3 HL Confirm whether you work by a motor signal check normally The chuck is open Check turning a tip with a fin...

Page 11: ...18 Please consult with dealer from whom you purchased it about waste disposal 16 Disposing Product ...

Reviews: