NRG INSTRUCTIONS
SRA System | Soiling Measurement Kit
SRA Soiling Measurement Kit Instructions
Rev. 6.0
[email protected] | Page 18
19 December 2018
Temperature Soiled:
Expand Channel 22 and choose “Custom Analog 0-1V” from the “Load From Defaults” drop-
down menu.
Choose “P-SCM #9135, (0 to 30) V, SE Input, Constant 12V EXC” from the SymphoniePRO Signal
Conditioning Module (P-SCM)” drop-down menu.
Edit the Description, Scale Factor, Offset and Units as show in the image below:
I
sc
Clean
:
Expand Channel 24 and choose “Custom Analog 0-5V” from the “Load From Defaults” drop-
down menu.
This will auto select “P-SCM #9132, (0 to 5) V, SE Input, Constant 12V EXC” from the
SymphoniePRO Signal Conditioning Module (P-SCM)” drop-down menu.
Edit the Description, Scale Factor, Offset and Units as show in the image below: