8. Technical Data
Headstage:
Input voltage range:
±1200 mV
Operating voltage:
±12 V
Enclosure:
Size: 40 x 70 x 20 mm, driven shield (COMMAND potential)
mounting plate: 50 x 70 mm, not conducting
Electrode connector:
BNC with shield (COMMAND potential)
Ground connector:
1 mm connector (black)
Command output:
1 mm connector (red)
Current range:
±20 nA max.
Feedback resistor:
500 M
Noise: <1
pA
LP FILTER:
Bessel Filter
Attenuation:
4-pole: -24 dB/octave; 8-pole (optional): -48 dB/octave
corner frequencies:
20, 50, 100, 200, 300, 500, 700, 1 k, 1.3k, 2k, 3k, 5k, 8k, 10k,
13k, 20k Hz
GAIN:
7-position rotary switch (x0.5, x1, x2.5, x5, x10, x25 and
x50 mV / pA)
Voltage source:
COMMAND VOLTAGE set by 10-turn potentiometer, range:
1000 mV, polarity selectable with polarity switch (+, 0, -),
display XXXX mV
OUTPUTS:
Impedance: 50
Linear voltage range:
±10 V
FILTERED:
BNC connector, sensitivity selectable with GAIN switch,
FILTERED with LP Bessel FILTER
UNFILTERED:
BNC connector, sensitivity selectable with GAIN switch
INPUTS:
Resistance: 1
M
COMMAND VOLTAGE: INPUT BNC connector, sensitivity: /10, also displayed
MONITORING:
Impedance: 50
GAIN:
BNC connector, 1 V per step, +1 V…+7 V
FILTER:
BNC connector, 1 V per step, -8 V…+7 V
COMMAND:
BNC connector, sensitivity: x10
Dimensions:
19” rackmount cabinet, 19” (483 mm), 10” (250 mm), 1.75”
(44 mm)
Power requirements:
115/230 V AC, 60/50 Hz, fuse 0.4/0.2 A, slow