If you intend to read the signal from the VA-10X into a data acquisition system:
connect a BNC cable from the data acquisition system to the FILTERED or
UNFILTERED connector (#
14
or #
15
, Figure 1). Additionally, you can monitor the
GAIN setting by connecting a BNC cable from the data acquisition system to GAIN
MONITOR (#
13
, Figure 1) and / or the FILTER setting by connecting a BNC cable
from the data acquisition system to FILTER MONITOR (#
12
, Figure 1)
If you intend to use an external voltage source (e.g. for cyclic voltammetry):
connect your external voltage source to the INPUT ÷10 mV connector (#
10
, Figure 1).
Remember that the input voltage will be scaled down by a factor of 10 at the
headstage. Note that, when an external voltage source is used, the 3-position toggle
switch controlling the internal voltage source should be set to “0”, unless you want to
sum the external voltage with the internal voltage source.