The actual temperature at the sensor now approaches the desired temperature but will not 
reach 37 °C because the control loop is not yet well adjusted. The actual temperature should 
arrest at 1 to 2 °C below the desired temperature. Please wait until this steady state value is 
reached. This may last several minutes. If this value is not reached set the GAIN to higher 
levels until - in this example – 35 °C to 35.5 °C is reached. If you don’t get a steady state 
value, set the LIMITER to 80-90% (or 30-40%, respectively) and wait again. 

Now turn on the INTEGRATION and set it to a low level 

INTEGRATION

OFF

MAX.

 
The actual temperature at the sensor will now reach the desired temperature possibly after a 
small overshoot. Wait again until the steady state value is reached. 

• 

If the desired temperature is not reached first set the GAIN to a slightly higher level and 
then – if necessary – raise the INTEGRATION 

 

 

Note

:  The speed of the control circuit (and therefore the time in which the actual 

temperature reaches the DESIRED TEMPERATURE) is highly dependent on the first 
settings of GAIN and INTEGRATION. Thus, when beginning to use the MTC-20/2S we 
recommend to try out several first settings of GAIN and INTEGRATION to find out which 
settings are best for your experimental conditions. 
 
If you use a perfusion system it is recommended to heat the solution approximately to the 
desired temperature 

before

 perfusing the bath either with a heating foil wrapped around the 

supplying tube (available from npi) or by heating up the supplying bowl.  
 
As mentioned above it is very important that the sensor element and the heating element are 
in good thermal connection. 
 

___________________________________________________________________________ 
version 2.7 

 

 

page 14 

Summary of Contents for MTC-20/2S

Page 1: ...DESCRIPTION FOR THE MTC 20 2S TEMPERATURE CONTROL SYSTEM VERSION 2 7 npi 2006 npi electronic GmbH Hauptstrasse 96 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support npielectron...

Page 2: ...ntents 1 Safety Regulations 3 2 MTC 20 2S Components 4 3 MTC 20 2S System 4 3 1 System Description 4 3 2 Description of the Front Panel 5 3 3 Rear Panel Elements 8 3 4 Electronic Control and Protectio...

Page 3: ...riate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Before opening the cabi...

Page 4: ...ystem guarantees low noise operation and has special protection features to prevent the preparation from damage The MTC 20 2S incorporates two electronic thermometers for small semiconductor sensors R...

Page 5: ...e type of the element in lowercase letters Then a short description of the element is given 1 DESIRED TEMP C potentiometer Potentiometer to set the working temperature of the controller system If the...

Page 6: ...to 5 INTEGRATION potentiometer Potentiometer to set the INTEGRATION parameter time constant of the temperature controller PI controller time range 50 ms 20 s 6 Display B Digital display to indicate t...

Page 7: ...the temperature determined by SENSOR A 13 sensitivity 10mV C 13 SENSOR A connector 8 pole connector for temperature SENSOR A This connector also provides the POWER OUTPUT 18 19 for the heating unit ma...

Page 8: ...onnectors Banana jack connectors for the heating unit This output supplies a maximum voltage of 12V DC and a maximum current of 1 A POWER OUTPUT is also available at the SENSOR A 13 and SENSOR B conne...

Page 9: ...control must be in good thermal connection with the heating wire If not overheating and damage may occur ALARM circuit The internal thermometer R2252 sensor inputs are protected by the sensor alarm ci...

Page 10: ...bled Modes of Operation Overview DIRECT MODE output voltage is set directly by the DESIRED TEMP C control Internal SENSOR mode output voltage is controlled by the PI controller using SENSOR A or SENSO...

Page 11: ...e the cabinet is always connected to protective earth green yellow wire The system can be grounded either via a BNC cable connected to INPUT or OUTPUT 10 mV C or via the blue OUTPUT GND connector 5 Co...

Page 12: ...__________________________________________________________________________________________________ Figure 4 Block diagram of the MTC 20 2S _____________________________________________________________...

Page 13: ...tage Please follow these instructions step by step to avoid problems in adjusting the instrument Turn all potentiometers to the most left position switch the MODE selector to the position SENSOR A or...

Page 14: ...rshoot Wait again until the steady state value is reached If the desired temperature is not reached first set the GAIN to a slightly higher level and then if necessary raise the INTEGRATION Note The s...

Page 15: ...pped with foam Starting temperature was 27 3 C Starting conditions were LIMITER set to 0 GAIN set to 0 INTEGRATION off 25 27 29 31 33 35 37 39 0 100 200 300 400 500 600 700 800 900 1000 time s tempera...

Page 16: ...s below 3 C not connected or broken sensor or above 45 C short circuited sensor Digital displays 3 1 2 digits XX X C temperature of SENSOR A or B or XX X V voltage at POWER OUTPUT COMMAND INPUT BNC co...