![Now Technologies GyroSet Cubo User Manual Download Page 35](http://html1.mh-extra.com/html/now-technologies/gyroset-cubo/gyroset-cubo_user-manual_1713558035.webp)
SPECIAL SAFETY CHECKLIST
Please make sure that the following criteria are met before trying to drive the wheelchair
with the GyroSet Drive System:
❏
Understanding of the intended use of the GyroSet Cubo and the GyroSet Vigo by getting
familiar with this User Manual.
❏
The adjustment of the headrest: The user must be able to reach the Headrest Sensor at
any time to be able to activate it.
❏
The intended direction of movement must be set by using Glory Tools.
❏
The user’s head movement range must be set by using Glory Tools.
❏
Out of operation range for the faint detection threshold must be set using Glory Tools.
❏
Tremor filter must be set using Glory Tools.
❏
Click sensor functions and parameters must be set using Glory Tools.
❏
Click sensor long-press maximal timeout must be set using Glory Tools.
❏
Centre mode and parameters must be set using Glory Tools:
❏
Please make sure to check if the center mode is set to “Countdown” in Glory
Tools. If yes, the “Timeout” has to be set.
❏
The maximum speed, acceleration, deceleration of the head drive profile of the
wheelchair controller must be adjusted:
❏
Settings of the controller must be administered specifically trained and authorized
personnel.
❏
Acceleration and deceleration settings of the wheelchair shall not cause inertia
on the user’s head that the user cannot compensate.
❏
Maximum speed must always be in accordance with the local regulations.
❏
The service network must be used only for the time when changes are made to the
personal drive profile.
❏
The user should avoid:
❏
Environments with extreme radio interference when driving wirelessly as it may
cause the wheelchair to stop if it receives no signals.
❏
The user shall be aware of the fact that the GyroSet Link must be selected by using the
Headrest Sensor for control with the GyroSet Vigo.
❏
The user must be trained and know how to handle the center detection (jump to center).
❏
In case the mode is set to countdown, the user’s head shall be motionless in a
neutral position after activating the drive function for a few seconds until the
countdown stops.
❏
In case the mode is set to automatic, the drive won’t start until the head is put
into a motionless state for less than a second.
❏
Be aware that the GyroSet Vigo acts as a joystick within the R-Net control system: