4.2 Outputs
Each output is fault and electrostatic discharge protected. Each output is
independent and any output can be faulted for an indefinite period of time with no
permanent damage. Each output is connected to a monitor circuit that detects a
local fault on the output. The fault status is indicated on the front panel. The fault
status and the protection on each output facilitates installation to help prevent
damage.
The nominal application is for a 10 MHz sine wave output in the range of 1 Vrms.
Skewing between channels < 5 ns.
4.3 Built in Test
There are number of power supplies in the design to meet special needs and noise
reduction. All power supply voltages are monitored, and can be accessed via
RS232. In addition, all current channel statuses, or Vrms values, can be
monitored, as well as power supply health.
4.4 Power Supplies
The unit is designed to accept power in the range of 90 to 264VAC, 50 to 60 Hz.
This allows global application. The design is such that no action need be taken to
operate from global power types. This feature avoids installation damage that
occurs in designs that require an input power switch mode be used.
There is an EMI filter between the internal power supply and the available power
being used. This filter minimizes the electrical noise from entering the circuitry
and negatively impacting noise performance. Also, in most applications, the
equipment that surrounds this unit is sensitive and the filter also reduces noise that
could impact the performance of other equipment.
If the optional DC power option is installed the unit can be powered from nominal
24 VDC or optional -60 to +60Vdc. The output of the DC to DC converter is
effectively diode OR’d with the AC supply and the DC power supply becomes the
prime when the AC power fails.