background image

LoopView-i 

NOVUS AUTOMATION 

2/3 

MECHANICAL INSTALLATION 

 

Fig. 03 

– Dimensions

 

OPERATION 

The controller’s front panel, with its parts, can be seen in the 

Fig. 04. 

 

 

Fig. 04

 – Indicator overview 

Display

: Shows the measured variable, the configuration parameters 

and their values / conditions. 

 Decrement Key and 

 Increment Key

: Allow altering the 

values of the parameters and accessing the indicator operation and 

configuration cycles.   

STARTUP 

When the power is on, the indicator shows the value of the measured 

process variable 

(PV)

. This is the 

Indication Screen

In order for the indicator to operate properly, its parameters need to 

be configured first. The configuration consists of the calibration and 

definition of the programmable interval. The user must be aware of 

the importance of each parameter and for each one determine a valid 

condition. 
The 2 cycles of parameters are: 

 

Operation 

 

Configuration 

When pressing both keys (

 

and 

)  at the same time, the 

indicator shows a screen where the device password must be 

entered. After entering the password, the indicator will show the 

configuration cycle. To advance the parameters of this cycle, use the

 

key with short touches. To rewind parameters, use the 

 

key. To select the parameter, press both keys (

 

and 

) at the 

same time. 

ACCESS PASSWORD 

On the login screen, “

0000

” will show on the display, to enter a 

value, press the 

, key, to change decimal places, press the 

 

key, and to confirm the password, press both keys (

 

and 

) at 

the same time, the password is requested “

66

”. 

PARAMETERS 

OPERATION CYCLE 

PV 

PV Indication Screen

. Value of measured variable. 

CONFIGURATION CYCLE 

Ad-L 

AD-L

 

Statement of the minimum calibration signal at the 

beginning of the range applied to the input (Low). 

Ad-h 

AD-H

 

Statement of the maximum calibration signal at the 

end of the range applied at the input (High). 

Ds-l 

dS-L

 

Sets the lower value of the indication range (Low). 

Ds-h 

dS-H

 

Sets the upper value of the indication range (High). 

-Dp- 

dP

 

Sets the number of decimal places that will be shown 

on the display. 

Line 

Line

 

Parameter that allows the user to make corrections to 

the indicated PV value. 

Filt 

Filter

 

Digital Input Filter - Used to improve measured signal 

(PV) stability. Adjustable between 0 and 9. At 0, it 

means minimum filter and at 9, maximum filter. The 

larger the filter, the slower the measured value 

response. 

end 

End

 

Final parameter that, when pressing both keys (

 

and 

)

 

at the same time, save the settings that 

have been set by the user. 

CALIBRATION 

Following is the 

LoopView

  calibration procedure, which can be 

divided into two steps, the 

unlocking

 and the 

calibration

 steps: 

Unlock:

 

1.

 

Connect 

LoopView

 into a 4-20 mA signal simulator. Vary it and 

check if the indicator is powered up accordingly between 4 and 

20 mA. 

2.

 

Press 

 and 

 keys simultaneously (from now on referred 

to as 

  / 

) for 3 seconds and release them. Then 

parameter 

Lccy

  will be displayed. It means we’ve accessed 

Configuration Cycle. 

3.

 

At 

Lccy

  parameter click 

  / 

  in order to enter this 

parameter’s adjustment. The current displayed value will be 

0000

. The 

 key increments values from 

0

 to 

9

 in a single 

digit. The 

 key shifts to the next digit to be adjusted. Define 

value 

0066

 at it and press 

 / 

 to confirm. 

Calibration:

 

4.

 

Parameter 

AD-L

 will be displayed. Click 

 / 

 to enter this 

parameter. A random number is displayed. 

5.

 

Simulate 4 mA in its input. 

6.

 

Single click 

 key. At this moment 

LoopView

 reads input’s 

current and determines it as its lower point (4 mA). Press 

 / 

 to confirm and go back to 

AD-L

7.

 

Press 

 key to shift to the next parameter, 

AD-k

8.

 

Press 

  / 

  to enter 

AD-k

. A random number will be 

displayed. 

9.

 

Simulate 20 mA in its input. 

10.

 

Click 

 key. At this moment 

LoopView

 reads input’s current 

and determines it as its superior point (20 mA). Press 

  / 

  to confirm and go back to 

AD-k

. Calibration is finished 

here. 

11.

 

Press 

 key until parameter 

END

 be displayed. 

12.

 

Press 

 / 

 to leave Configuration Cycle. 