2.4. The device produced with a basic configuration, focused on
heavy concrete middle marks. For other marks and materials
required calibration and adjustment in accordance with GOST
17624, GOST 24332 and methodological recommendations MDS
62-2.01 control the strength of concrete monolithic structures by
ultrasonic method of surface sounding.
2.5. The device provides a visualization of the received ultrasonic
waves (A signal), has a mode of the oscilloscope for viewing and
analyzing the signals A.
2.6 The device provides operation:
- at the surface sounding with the surface sounding probe on the
fixed base (120 ± 1) mm with a dry contact;
- at the through-sounding with ultrasonic probes at random basis
with contact lubricant or surface and angular sounding with dry
contact (protectors, conical nozzle) or with a lubricant on an random
basis.
2.7. Operating conditions: temperature range: from -
20 ° C to 50 ° C,
relative humidity up to 80% non-condensing humidity, atmospheric
pressure 86 ... 106 kPa.
3. Specifications
Measuring range of ultrasonic pulse spreading
time,
μs
10…999,9
Measurement resolution of time of ultrasonic
pulses,
μs
0,01
The limits of permissible basic error spreading time
of ultrasonic pulses,
μs
±(0,01t + 0,1),
where t
–
measurement
time