background image

Victoza

®

 Pen

Needle (example)

Flow check 

symbol

Dose  

button

Pointer

Dose 

display

Cartridge 

scale

Pen cap

Rubber  

stopper

Cartridge tip

Cartridge

Dose selector

Outer 

needle cap

Inner 

needle cap Needle

Protective 

tab

Take your new Victoza

®

 

pen out of the refrigerator. 
Wash hands with soap 
and water before use. 
Check pen label before 
each use to make sure it  
is your Victoza

®

 pen. 

Pull off pen cap. 
Check Victoza

®

 in the 

cartridge. The liquid 
should be clear, colorless, 
and free of particles.  
If not, do not use.
Wipe the rubber stopper 
with an alcohol swab. 

Remove protective tab 
from outer needle cap.
Push outer needle cap 
containing the needle 
straight onto the pen,  
then screw needle on  
until secure.

Pull off outer needle cap.  
Do not throw away. 

Pull off inner needle cap  
and throw away. A small  
drop of liquid may appear.  
This is normal.

Preparing Your Pen

 

Flow  

check 

symbol 

selected

0.6 mg  

selected

1.2 mg  

selected

1.8 mg  

selected

If you have questions about using your Victoza

®

 pen, call toll-free 1-877-484-2869 or visit victoza.com.

Please see Important Safety Information, including Boxed Warning, and Prescribing Information and 
Medication Guide, enclosed or at www.Victoza.com.

First-Time Use for Each New Pen

 

Hold pen with needle 
pointing up.
Tap cartridge gently  
with your finger a few 
times to bring any air 
bubbles to the top of  
the cartridge.

Keep needle pointing up 
and press dose button 
until 0 mg lines up with 
pointer. Repeat this up to 
6 times, until a drop of 
Victoza

®

 appears at the 

needle tip. 
If you still see no drop of 
Victoza

®

, use a new pen 

and contact Novo Nordisk 
at 1-877-484-2869. 

After setting up a new  

pen, follow these steps

Turn dose selector until flow 
check symbol (--) lines up 
with pointer. The flow check 
symbol does not administer 
the dose as prescribed by 
your health care provider.  
You only need to do this 
ONCE for each new pen and 
it is ONLY required the first 
time you use a new pen.

If this is the 

first time 

with a new 

pen, follow 

these steps

If you have 

already used 

your pen, follow 

these steps

Routine Use.

 

Prepare your Victoza

®

 pen for injection 

as described in the “Preparing Your Pen” section. The 
“First-Time Use for Each New Pen” section should only 
be performed when a new pen is used for the first time.

Routine Use and Injecting the Dose

Injecting the Dose.

 Insert needle into your skin in the stomach, thigh, or upper arm. Use the injection 

technique shown to you by your health care provider. 

Do not inject Victoza

®

 into a vein or muscle. 

Dial the Dose. 

Victoza

®

 pen can give a dose of 0.6 mg 

(starting dose), 1.2 mg, or 1.8 mg. Be sure that you 
know the dose of Victoza

®

 that is prescribed for you. 

Turn the dose selector until your needed dose lines up 
with the pointer (0.6 mg, 1.2 mg, or 1.8 mg). You will 
hear a “click” every time you turn the dose selector.

 

Do not set the dose by counting the number of 
clicks you hear. 

If you select a wrong dose, change it by turning the 
dose selector backwards or forwards until the correct 
dose lines up with the pointer. Be careful not to press 
the dose button when turning the dose selector. This 
may cause Victoza

®

 to come out. 

Withdraw Needle

You may see a drop of Victoza

®

  

at the needle tip. This is normal 

and it does not affect the dose  

you just received. If blood  

appears after you take the  

needle out of your skin, apply 

light pressure, but

 

do not rub 

the area

.

Carefully put the outer needle  

cap over the needle. Unscrew  

the needle. Safely remove the 

needle from your Victoza

®

 pen  

after each use.
Place used needles in a closeable, 

puncture-resistant container. If  

your Victoza

®

 pen is empty or if  

you have been using it for 30  

days (even if it is not empty),  

throw away the used pen. You  

may use a sharps container (such  

as a red biohazard container), a 

hard plastic container (such as  

an empty detergent bottle), or 

metal container with a screw top 

(such as an empty coffee can).
Ask your health care provider  

for instructions on the right way  

to dispose of your used needles,  

pens, and the container. Do not 

throw the disposal container  

in the household trash. Do  

not recycle.

0

mg

6

Keep the dose button pressed down and make sure that you keep  

the needle under the skin for a full count of 6 seconds to make sure  

the full dose is injected. Keep your thumb on the injection button  

until you remove the needle from your skin. 
Rotate your injection sites within the area you choose for each dose.  

Do not

 use the same injection site for each injection.

Press down on the center  

of the dose button to inject until  

0 mg lines up with the pointer.
Be careful not to touch the dose 

display with your other fingers. 

This may block the injection. 