Novis ProCair Quad User Manual Download Page 14

14

User Guide

 ProCair Quad

Operation

CPR Release

Rapid deflation of the mattress may be required for 
emergency treatment (or to decommission the unit).

The Rapid CPR valve is located at the top of the 
mattress, to the right of the patient’s head.

Mode

In Alternating Mode, alternate mattress cells 
inflate and deflate following a fixed cycle time 

of 15 minutes, with the exception of static head cells. 

Alternating mode is used for normal therapeutic 
function. 

In Static Mode, all mattress cells remain fully 
inflated thereby suspending the alternation 

cycle and any therapeutic patient benefits. 

This mode should be used to create a firm base for 
stable patient handling and transport or other special 
circumstances.

  The system will automatically start in Static 

Mode to initiate mattress inflation, and revert 
to Alternating Mode once optimal pressure is 
reached.

  The system will operate in Static Mode for a 

maximum of 20 minutes, after which it will 
automatically revert to Alternating Mode for 
patient safety. 

Auto Fowler Boost

Fowler Boost mode will automatically 
become active whenever the head of the 
mattress is inclined to a 15° or greater 
angle.

This mode increases air pressure to the cells, to 
compensate for the additional load of a seated 
patient. This automatic safety measure allows 
patients to remain in a seated position while 
minimising the risk of bottoming out.

Once the head of the mattress is brought back to a 
reclined position (or below a 5° angle), Fowler Boost 
mode will automatically disengage.

FOWLER

ALTERNATE

STATIC

Backup Battery

The lithium-ion backup battery is designed to provide 
temporary uninterrupted therapy when mains power 
is unavailable eg during patient transfer or in a power 
outage. 

The control unit automatically switches to battery 
power when mains power is disconnected during 
operation (when system is switched on and running). 

An alarm will sound to alert users that the control 
unit has switched to battery operation in case the 
change was unintended, so that mains power can be 
reconnected if necessary. Press MUTE to disable the 
audible alarm, which will not sound again until mains 
power has been restored and disconnected again.

Up to 3 hours of continuous operation can be 
expected from a new, fully charged battery, however 
battery capacity will decrease with use as it ages.

The backup battery automatically recharges when 
mains power is connected. The control unit does not 
need to be switched on to recharge the battery.

Recharge Time:

Control Unit OFF - up to 6 hours

Control Unit RUNNING - up to 24 hours

the BATTERY CHARGE INDICATOR light will flash in 
green during charging.

When the system is not in use for extended periods, it 
is recommended that the battery be charged every 6 
months to maintain its condition.

The battery unit is a consumable part and has a 
usable life of up to 3 years and should be replaced by 
an authorised Novis repair agent.

Summary of Contents for ProCair Quad

Page 1: ...1 ProCairQuad User Guide U S E R G U I D E ProCairQuad...

Page 2: ...provement and reserves the right to amend specifications presented in this guide Information correct at time of production February 2019 2019 Novis Healthcare All rights reserved Important Notice Defi...

Page 3: ...utions 6 7 System Preparation 8 10 Patient Set Up 11 Operation Control Unit 12 13 Operation Mode 14 Operation Pull Cap CPR Release 14 Operation Transport Mode 15 Operation Deflation Storage 15 Care an...

Page 4: ...ial and venous capillary occlusion in the patient s surface tissue maintaining and stimulating the flow of blood and lymphatic fluids through these tissues to provide essential oxygen and remove metab...

Page 5: ...vironment Temperature 5 C to 60 C 41 F to 140 F Humidity 30 to 90 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The ProCair Quad mattress...

Page 6: ...part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams objects in pockets etc Never use sharp objects or electrical...

Page 7: ...e of smoking materials or open flame air flowing through the mattress will support combustion Do not open the control unit risk of electrical shock Refer servicing to qualified service personnel Dispo...

Page 8: ...ase Ensure the buckles are securely fastened and straps are pulled tight Do not secure mattress straps to bed side rails straps will tear Ensure that straps do not interfere with the operation of the...

Page 9: ...ith other control unit which looks similar Straighten any twists in the umbilical cord air hoses to ensure uninterrupted air flow between the control unit and mattress Ensure the umbilical cord is not...

Page 10: ...alarmguideonleft ALARMINDICATORS BATT IN USE 9When initial inflation is complete the Pressure Setting and Max Inflate indicators will extinguish to indicate the system is ready for use The system auto...

Page 11: ...the cover to feel the cells for inflation 2With the patient lying supine unzip one side of the top cover just past sacral region lower spine 3Slide your hand underneath the patient and feel for a def...

Page 12: ...de on the control panel to look up the fault indicated also refer to the corresponding troubleshooting guide page 19 for detail Alarms include LOW PRESSURE possible loss of pressure HIGH PRESSURE matt...

Page 13: ...ut not switched on When periodical servicing is required the amber Standby light will flash Please contact Novis Healthcare for system maintenance This equipment must only be serviced by a qualified s...

Page 14: ...g the risk of bottoming out Once the head of the mattress is brought back to a reclined position or below a 5 angle Fowler Boost mode will automatically disengage FOWLER ALTERNATE STATIC Backup Batter...

Page 15: ...sed Heel Therapy The specialised HeelCair zone enables the seven narrow alternating air cells in the heel zone to be disconnected to remove all pressure on a patient s heels without interruption to ma...

Page 16: ...T IN USE Deflation and Storage 1 Press the power button for a minimum of three seconds to switch off the control unit 2Switch off mains power and unplug the power cord from the mains outlet 3Open CPR...

Page 17: ...se and Cells Cleaning Swab the mattress base and cells with a solution of sodium hypochlorite or similar up to 10 000 ppm available chlorine Dry thoroughly before reassembly Do not machine wash or tum...

Page 18: ...the zipper running towards the CPR Then close both zippers by zipping towards the foot end of the mattress Air Filter Replacement 1 Switch off power supply to the control unit and disconnect the air h...

Page 19: ...power after 1 minute and restart initiation Remove patient from mattress Turn off control unit and disconnect process return patient to mattress once initiated and alternation has resumed If issue per...

Page 20: ...power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if necessary Checkconditionofpowercordandplug Checkifmainssocketisfault...

Page 21: ...T 0 2 A MAX FUSE RATING T1AH 250V MIN MAX PRESSURE 20 60 mmHg 6 mmHg PROTECTION TYPE Class II Type BF INGRESS PROTECTION RATING IP21 MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEI...

Page 22: ...onsible and contact your local authority on available options to recycle this product at its end of life Service Life The expected service life of a control unit and a mattress is highly dependent on...

Page 23: ...t a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum ou...

Page 24: ...60 Hz 50 Hz Conducted RF IEC 61000 4 6 3 Vrms 0 15 MHz 80 MHz 3 Vrms 0 15 MHz 80 MHz Portable and mobile RF communications equipment should be used no closer to any part of the ProCair Quad control u...

Page 25: ...Hz 0 2 0 3 9 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation B 18 Hz 2 0 3 28 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25...

Page 26: ...may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for any repair other than th...

Page 27: ...27 ProCairQuad User Guide Notes...

Page 28: ...Pressure care and patient handling specialists novis com au 1300 738 885 V1 June 2019...