10
7. Dispensing
oils
To use the aromatic oils you must do the following.
The dispenser is located on the back of the device (zdj. 3), pull it out (zdj. 4), then on the
sponge in the middle (zdj. 5), put the right amount of oil (as desired). Then, close the
dispenser to the position as in a photo (zdj.6). The device is ready for use.
8. Cleaning and Maintenance
8.1 Water Tank
Before cleaning, ensure that the unit is switched off. Clean after every use and do not
keep water in the tank for longer than 2 days.
8.2 Transducer
To clean the (main body) transducer, add some de-scaling cleanser (about 0.5 g
powder) onto the transducer and then add a small amount of water and wait for 2-5
minutes. Use a soft brush to clean. Do not use soap or other alkaline substances to clean
the transducer.