LASER SAFETY
PAGE
|
47
Laser Safety
Table 2-1 Class 4 safety features.
1
Not available on OEM lasers
When integrating Novanta OEM lasers, the Buyer and/or integrator of the end system is responsible for
meeting all applicable Standards to obtain the CE mark. To aid this compliance process, Novanta testing
program has demonstrated that lasers comply with the relevant requirements of 2014/30/EU, the
Electromagnetic Compatibility Directive, as summarized in the table below.
Feature
Location/Description
Required by:
CDRH EN60825-1
Available on
OEM i401
Keyswitch
1
Rear panel control
On/Off/Reset Key switch controls power to laser
electronics. Key cannot be removed from switch in
the “On” position.
Yes
Yes
No
Shutter Function
Laser Control
Functions as a beam attenuator to disable RF
driver/laser output when closed.
Yes
Yes
Yes
Shutter Indicator
Rear Panel Indicator (Blue)
Illuminates blue to indicate shutter is open.
No
No
Yes
Ready Indicator
Rear Panel Indicator
(Yellow)
Indicates that laser has power applied and is capable
of lasing.
Yes
Yes
Yes
Laser Indicator
Rear panel indicator (Red)
Indicates that the laser is actively lasing. Lase LED
illuminates when the duty cycle of the Command
signal is long enough to produce laser output.
No
No
Yes
Five second delay
Circuit Element
Disables RF driver/laser output for five seconds after
Key switch is turned to “On” or remote reset/start
pulse is applied when Key switch is in “On” position.
Yes
No
Yes
Power Fail Lockout
1
Circuit Element
Disables RF driver/laser output if input power is
removed then later reapplied (AC power failure or
remote interlock actuation) while Key switch is in
“On” position
Yes
Yes
No
Remote Interlock
Side Panel Connection
Disables RF driver/laser output when a remote
interlock switch on an equipment door or panel is
opened.
Yes
Yes
Yes
Remote Interlock
Indicator
Rear panel indicator (Green/Red)
Illuminates green when Remote Interlock circuitry is
closed. Illuminates red when interlock circuitry is
open.
No
No
Yes
Over Temperature
Protection
Circuit Element
Temperature shutdown occurs if temperature of the
laser tube rises above safe operating limits
No
No
Yes
Temp Indicator
Rear panel indicator (Green/Red).
Illuminates green when laser temperature is within
operating limits, changing to red when thermal limits
are exceeded.
No
No
Yes
Warning Labels
Laser Exterior
Warnings to personnel of potential laser hazards.
Yes
Yes
Yes