This section describes how to perform Linearity tests with the
StatSensor Xpress-i Creatinine Meter. There are 5 levels in the
StatSensor Creatinine Linearity kit.
Refer to the StatSensor Creatinine Linearity Kit
package insert sheet for complete instructions, indi-
cations, precautions, and limitations of the system.
1. Insert a test strip into the meter. All segments of screen will
display for 2 seconds. Then a flashing blood drop will display.
NOTE:
If strip is re-
moved before the test
starts or is not used
for over 2 minutes,
the screen will go
blank.
Testing a Linearity/Proficiency Solution