TCM Endo R 11 

 

Operating Instructions, TCM Endo R 11, Ref. 31649, V48/15

 

13 

 

EN

10

 

Malfunction and troubleshooting

 

Malfunction Cause  Solution 

Reference 

in 

manual 

Device is not operational 

Control unit is not  
switched on 

Switch main switch “I/O”  
to “I” 

7.1 Switching device on and off 

No connection to  
electricity- supply 

Connect power cord to  
electricity-supply 

6.3 Device preparation 

Wrong voltage 

Check power supply and 
change voltage if needed 

6.2 Connection to the  

power supply 

Defective fuses 

Replace fuse 

9.1 Replacing the control unit fuse 

Pedal is not functioning 

Control unit is not  
switched on 

Switch main switch “I/O”  
to “I” 

7.1 Switching device on and off

 

Pedal is not connected  
properly 

Check plug connection,  
replug 

6.3 Device preparation

 

Micro motor is not  
functioning 

Control unit is not  
switched on 

Switch main switch “I/O”  
to “I” 

7.1 Switching device on and off 

Motor is not connected  
properly 

Check plug connection,  
replug 

6.3 Device preparation 

Contra angle is not work-
ing properly 

Contra angle is not connected 
properly to the motor 

Press contra angle firmly to 
motor, check seating 

6.3 Device preparation 

 

If a fault cannot be rectified, please contact your supplier or an authorized service center. The addresses are provided on 
the last page of tis operating instructions. 
 
 
 

11

 

Accessories and spare parts 

 
Spare parts 

Ref. 

Mikro motor Endo ---------------------------------------------------------------------------------------------------------------------------- 1902 
Contra angle 2:1 ------------------------------------------------------------------------------------------------------------------------------- 1683 
On/Off pedal ---------------------------------------------------------------------------------------------------------------------------------- 1479 
 

Accessories 

Ref. 

Transport suitcase TCM Endo R 11 -------------------------------------------------------------------------------------------------------- 1078 
 

To order additional parts, please contact our customer service department. 
 
 

12

 

Information on disposal 

When disposing of the device, device parts and accessories, the regulations prescribed by law must be observed. 

 

Do not dispose of devices with houshold waste! 
To ensure environmental protection, old devices can be returned to the dealer or manufacturer. 
 
 

Motors that have reached the end of their service life may not be disposed of with houshold waste. 
Motors must be sterilized before disposal. Please observe currently valid national disposal regula-
tions for infectious waste. 

 
 

Summary of Contents for TCM Endo R11

Page 1: ... further development FR Sous réserve de modifications techniques En raison du développement constant de nos produits les figures et les caractéristiques techniques peuvent différer légèrement ES Reservado el derecho a realizar modificaciones técnicas Las imágenes y los datos técnicos pueden variar mínimamente debido al desarrollo del producto PT Resevamo nos o direito a alterações técnicas O desen...

Page 2: ... overview 5 6 Startup 6 6 1 Device setup 6 6 2 Connection to the power supply 6 6 3 Device preparation 7 7 Operation 8 7 1 Switching device on and off 8 7 2 Operation panel 8 7 3 Setting up parameters 8 7 4 Starting motor 9 7 5 Automatic Limiter AL 9 7 6 Automatic Reverse AR 9 7 7 Reciprocal mode 9 7 8 Function control 9 8 Cleaning disinfection and sterilization 10 8 1 Control unit and On Off peda...

Page 3: ... general patient status or the risks of anesthesia related to the treatment 1 3 Technical data TCM Endo R 11 1 4 Ambient conditions Transport and storage Operation Relative humidity Max 90 Max 80 Temperature 0 C 60 C 10 C 30 C Atmospheric pressure 700 hPa 1060 hPa 800 hPa 1060 hPa 1 5 Warranty coverage Purchasing the TCM Endo R 11 entitles you to a 1 year warranty If you return the warranty card f...

Page 4: ...e BF applied part is the connected instrument Electrical and electronic devices that have reached the end of their service life comprise hazardous waste and may not be disposed of together with household waste Valid local disposal regulations apply Certified by the Canadian Standards Association CSA Symbol indicating the serial number with the date of manufacture year month CE symbol with notified...

Page 5: ...xtures At the choosing of the instrument the operator has to make sure it s biocompatible according to EN ISO 10993 Electric micromotors generate significantly more power than traditional air turbines and air motors Poorly maintained worn damaged or misused handpieces may generate frictional heat capable of causing serious burn injuries to the patient Heat generation may occur quickly and without ...

Page 6: ...t for On Off pedal 6 Plug of On Off pedal 7 2 1 Endo Contra Angle 8 Instrument for root cannal preparation optional 9 Handpiece cradle 10 Mikro motor Endo 11 Socket for Mikro motor Endo 12 Plug of Mikro motor Endo 13 Fuse compartment with voltage setting 14 Device main switch 15 Power plug socket 16 Type plate with type designation reference number serial number information on power supply and dev...

Page 7: ...e must be accessible at all times 6 2 Connection to the power supply Before plugging the power cable into the power socket for the first time you must check the supply volt age setting next to the power switch In order to prevent the risk of an electrical shock the device may only be connected to a power network with a PE protective ground conductor If the voltage shown does not correspond to the ...

Page 8: ...motor socket M of the control unit 3 Attache the contra angle with clamped instru ment to the micro motor Check seating of the contra angle with a gentel counter movement 4 Plug the power cord with the appropriate end into the socket of the control unit 5 Connect the power cord with the electrical outlet Before switching on make sure that the power supply of the device matches the country s specif...

Page 9: ...he device 7 2 Operation panel n d e r g e 7 3 Setting up parameters Reduce speed Increase speed Reduce torque Increase torque Select between AL and AR function Select rotation direc tion of instrument Activate reciprocal mode Select speed with the Speed und Speed keys Select torque with the Torque und Torque keys Switch between Auto Limiter and Auto Reverse mode Refer to chapter 7 6 Automatic Limi...

Page 10: ...ental debris for a fast and effective root cannal preparation 7 8 Function control Prior to the startup of the TCM Endo R 11 and the use of accessory equipment the user must always ensure that each individual component is in good working order free from defects clean sterile and operational All inscriptions on the device and its accessories must be readable and there must be no loose parts in the ...

Page 11: ...rilized material is not used immediately the material packaging must be labeled with the sterilization date Nouvag AG recommends including a sterility indicator 8 1 Control unit and On Off pedal Control unit and On Off pedal do not come into contact with the patient Wipe the outside using micro biologically tested surface disinfectant or a 70 isopropyl solution Don t use spray cleaning and spray d...

Page 12: ...nd disinfection spray contra angle thoroughly with Nou Clean spray and briefly wipe with a moistened cloth See illustrations on the spray can Packaging Single Pack the contra angle in an individual sterile supply package The bag must be large enough that the seal is not under tension Sets Arrange the contra angles into the foreseen trays or put onto an all purpose sterilization tray Sterilization ...

Page 13: ...slot 4 Fuse 1 5 Fuse 2 9 2 Safety inspections The performance of safety inspections on medical devices is required by law in several countries The safety inspection is a regular safety check that is compulsory for those operating medical devices The objective of this measure is to ensure that device defects and risks to patients users or third parties are identified in time An inspection interval ...

Page 14: ... connection replug 6 3 Device preparation Contra angle is not work ing properly Contra angle is not connected properly to the motor Press contra angle firmly to motor check seating 6 3 Device preparation If a fault cannot be rectified please contact your supplier or an authorized service center The addresses are provided on the last page of tis operating instructions 11 Accessories and spare parts...

Reviews: