ES3 "Intelligent" Pressure and Level Transmitter   

 

 

 

 

EX-1861

3.  Adjust the 

SPAN

 with the push buttons to 29 psi and push the        button until save appears 

on the display.

4. The transmitter will return to the home screen. 

The span can also be adjusted to a real process situation. The transmitter will measure the  
pressure in an actual process. This measurement will be used as the span value (20 mA).
1. Go to the home screen, the transmitter will display the actual measured value.
2.  Push the        button until the actual measured span and the message 

SPAN PUSHED

 appear 

on the display.

3. The transmitter will return to home screen.

 

P102 is the adjustment of the total span. 

 

When a compound range must be adjusted (for example -14.5 till +43.5 psi), a  

 

 

pan of 29 psi must be programmed. 

 

The Zero (P101) must be set at -14.5 psi. The transmitter is adjusted at - 14.5 psi =    

 

Zero and +43.5 psi = Span.

 

 

CANCEL MOUNTING POSITION EFFECT

 

 

 All transmitters are calibrated vertically. If the transmitter is installed horizontally, the 
transmitter has a small "mounting position" effect on the zero. The pressure value 
displayed, will be for example 0.003 psi instead of 0.000 psi. 

1. Navigate to program point 

P103

, and push the menu button to enter the menu.

2. Two choices appear on the screen: 

Set

 and 

Reset

.

 Choosing 

Set

 will adjust the zero to 0.000 psi in the mounting position when applicable.

  • 

Select 

Set

, and push the menu button to confirm. 

 

 

• The corrected is value is shown on the display.

 

 

• Push the menu button to save.

 Choosing 

Reset

 will put the transmitter back to factory setting. (vertical adjustment)

  • 

Select 

Reset

, and push the menu button to confirm to reset back to factory setting.

 

 

• The transmitter will return to the home screen.

 

 

Do not apply pressure when executing "Cancel mounting position effect"

 

 

DISPLAY SETTING OF UNITS

 

 

Various engineering units can be displayed on the display. 

  Factory setting = psi

1. Navigate to program point P104, and push the menu button to enter the menu.
2.  Several engineering units can be selected. Each selected engineering unit is automatically  

converted to the correct value of the corresponding unit. 

3. Navigate through this menu and choose the required unit, push the menu button confirm.
4.  The transmitter will return to the home screen. The measured reading will be displayed in the 

selected unit in the home screen.

  

The selected pressure unit is only visible on the display when UNITS is  

 

 

chosen in program point P109.

Summary of Contents for King-Gage 5800 Series

Page 1: ... recommendations and warnings in this manual before the instrument is installed For personal safety optimal use and maintenance of the KING GAGE 5800 Series ES3 Transmitter these instructions should be studied carefully KING GAGE A NOSHOK Company I 1010 West Bagley Road Berea Ohio 44017 I Ph 855 367 2494 I www king gage com ...

Page 2: ...or merchantability are given by this agreement The exclusive remedy for nonconformity of these goods shall be repair and or replacement of the nonconforming goods or parts Seller will not be liable for consequential damages resulting from breach of this agreement The term consequential damages shall include but shall not be limited to damage toallmachines equipmentandgoodsotherthanthegoods sold he...

Page 3: ...3 5 Wiring 4 4 Remaining 5 4 1 CE EMC Rules 5 4 2 Display and push buttons 5 4 3 Display read out 5 5 Summary Programming Points 6 5 1 Explanation Programming Points 6 P101 Zero Adjustment Zero 6 P102 Span Adjustment Span 6 P103 Cancel Mounting Position Effect 7 P104 Display Setting Of Units 7 P105 Output Selection 4 20 Ma Or 20 4 Ma 8 P106 Damping Adjustment 8 P107 Temperature Units 8 P108 Device...

Page 4: ... laser welded and helium leak tested A very small amount of oil fills the chamber surrounding the sensor and transfers pressure from the flush mounted dia phragm to the sensor Pressure on the sensor element creates a very small deflection of the silicon substrate and bridge network The resulting strain in the silicon resistors causes a change in the bridge resistance that is proportional to the pr...

Page 5: ...st be connected to the terminal board The wiring terminals can be operated without a screwdriver The opening levers of the terminal can be lifted and pressed down by hand Lift the opening levers of the termi nals and insert the corresponding wires Press down the levers by hand The terminal spring will close and the wire is clamped 3 5 WIRING The transmitter must always be connected to ground The t...

Page 6: ...ast display for optimal readout The menu is controlled by 3 push buttons Navigate with the up and down buttons through menus and measuring values Enter a menu and confirm selections with the menu button Push the up button to browse through various menus and adjusting values Push the down button to browse through various menus and adjusting values Push the menu button to enter menus or confirm a se...

Page 7: ...nt P104 2 Navigate to program point P101 and push the menu button to enter the menu 3 The actual measured value appears on the display 4 Increase this value with the arrow button to 1 45 psi push the button till save appear on the display 5 The transmitter will return to the home screen The measurement value at atmospheric pressure is now 1 45 psi With a applied pressure of 1 45 psi the transmitte...

Page 8: ...e 0 003 psi instead of 0 000 psi 1 Navigate to program point P103 and push the menu button to enter the menu 2 Two choices appear on the screen Set and Reset Choosing Set will adjust the zero to 0 000 psi in the mounting position when applicable Select Set and push the menu button to confirm The corrected is value is shown on the display Push the menu button to save Choosing Reset will put the tra...

Page 9: ...y setting The transmitter will return to the home screen TEMPERATURE UNITS In this menu the preferred temperature unit can be selected Factory setting Fahrenheit 1 Navigate to program point P107 and push the menu button to enter the menu 2 Two choices appear on the screen Fahrenheit and Celsius 3 Make a choice and push the menu button to confirm 4 The transmitter will return to the home screen DEV...

Page 10: ... Five choices appear on the screen Mode Cntrl Cmd number Message Period and Trigger 3 Select Mode Cntrl and push to confirm 4 Two choices appear on the screen On and Off Choose On to turn on burst mode Choose Off to turn off burst mode 5 Select Message to select the burst message 0 1 2 or 3 and push the menu button to confirm 6 Select Cmd number and push the menu button to confirm Five choices app...

Page 11: ...sing The Burst message is triggered when the measured value rises above the triggered value Falling The Burst message is triggered when the measured value falls below the triggered value On Change The Burst message is triggered when any value in the measuring changing Choose the desired burst mode and set the preferred parameters INFORMATION This menu summarizes information about the configuration...

Page 12: ...Page 12 ES3 Intelligent Pressure and Level Transmitter EX 1861 ...

Reviews: