Nor265 User Guide
Norsonic AS
27
Instruction
Command
Parameter Notes
Program Flash
memory with
running SW.
BFBurnFlash
Intended for use by Norsonic only.
Display memory
DM x
For debugging by Norsonic only.
8.2.3 LP
command
When the
LP command
is executed, a list of the parameters associated
with position 1 to 8 for the front panel switch is returned: The list has
the following format where the first number in each line corresponds
to the switch position number.
1, <Type>, <Info> <CR> <LF>
2, <Type>, <Info> <CR> <LF>
3, <Type>, <Info> <CR> <LF>
4, <Type>, <Info> <CR> <LF>
5, <Type>, <Info> <CR> <LF>
6, <Type>, <Info> <CR> <LF>
7, <Type>, <Info> <CR> <LF>
8, <Type>, <Info> <CR> <LF>
The <Type> of operation is indicated by a number between 0 and 5.
The <Info> will contain a list of parameters relevant for the Type. The
designators for the parameters correspond to the commands in the list
for remote commands. Each parameter is separated by a comma
symbol ",".
<Type>
Type
Parameters in <Info> field
0 Stop
[SP]
1 Sweep
[ST]
TA
TT
SA
SB
2
Go to [GT]
TA
TR
GT
3
Go relative [GR]
TA
TR
GR
4
Continuous positive rotation [CP]
TA
TR
5
Continuous negative rotation [CN]
TA
TR
Summary of Contents for Nor265