Nor265 User Guide
Norsonic AS
23
1
5
6
9
Male connector
outside view
8 Remote
operation
The turntable may be remotely controlled by a RS 232 serial interface,
(default 9600 baud) with hardware handshake. Commands other than
reading status are ignored if the user operation switch is not set to
remote.
8.1 Socket
Socket for remote command: 9 pin D-sub male, see figure.
If the computer or other controlling device has a 9-pin serial port with
a male connector, use a null modem cable like Nor1463 for the
connection.
8.2 Remote
commands
8.2.1 General
Remote operation is done most easily by using the accompanying MS-
Windows program
N o r 1 0 2 9 .
This allows easy control of
N o r 2 6 5
from WS-Windows based PC. The program may also be used for
programming the operation of the rotary switch on the front of
N o r 2 6 5
. If you are using the program, you may skip most of the
information in this section. However, if you want the operation of
N o r 2 6 5
to be smoothly integrated into your own program, you need
the detailed information in this section.
Pin no
Signal
Dir
1
2 RD
In
3 TD
Out
4
5 GND
6
7 RTS
Out
8 CTS
In
9
Housing GND