Norma Icon-5 Maintenance Manual Download Page 2

 

 

TDMG0002 Icon-5 Maintenance Guide 

– Version: 02 – Release date: 2020-04-27 

 

 

Maintenance procedure on Icon-5 analyzers

 

These steps must be performed by trained service engineer! Please note the following 
warnings:

 

 

WARNING 

Risk of personal injury. Movements of motors are controlled by limited 
current. Do not override built-in safety measures, do not short circuit fuses 

 

 

BIOHAZARD 

Risk of biological infection, contamination 

 

WARNING 

The analyzer should only be operated without covers with extreme 
caution, moving parts can cause injury. High Voltage board HVDCDC 
produces 180VDC. 

Service marked components only when the power source is disconnected. 

 

WARNING 

Risk of personal injury. The ceramic shear valve has a sharp needle 
installed. Avoid touching the needle 

 

To prepare the Icon-5 for the maintenance, please, shut down the analyzer and pinch 
the reagent pickup tubes using the clamps provided in the reagent pickup tube set, 
and remove both side covers to get easy access to the inner components. 

 

 

Summary of Contents for Icon-5

Page 1: ...ct our support team at support normadiagnostika com if further information or clarification is required Icon 5 Maintenance Kit HAT827A for 1 year 10 000 sample period Item Part pcs Aperture maintenanc...

Page 2: ...nfection contamination WARNING The analyzer should only be operated without covers with extreme caution moving parts can cause injury High Voltage board HVDCDC produces 180VDC Service marked component...

Page 3: ...sults in deepening cracks on the tube material which will eventually lead to leaks and reduction or failure in liquid movement The wash head will slowly fail to remove waste from the sampling needle w...

Page 4: ...s gently pull on the adapter to release it from the groove Pull the tubes off the tube adapters The kit contains two different sets of pump tubes Different tubes must be used for different pumps HAT80...

Page 5: ...on the tubes to make them sit properly in the pump over the rollers Reinstall the cassette Both ends must CLICK upon insertion Make sure that the cassette is level with the pump body Incorrect alignm...

Page 6: ...6 TDMG0002 Icon 5 Maintenance Guide Version 02 Release date 2020 04 27 Reinstall the pump Watch out for the external silicone tubes Avoid pinching the tubes and cables between metal plates...

Page 7: ...e keeps the measuring apertures free of unwanted air and regulates liquid flow during different stages of the measurement process If the rubber valve gets contaminated then it will not be able to clos...

Page 8: ...rve when the rubber valve base lays flat on the base of its socket Overtightening the screw cap will deform the valve and make its operation impossible Reinstall the aperture module The two pins must...

Page 9: ...9 TDMG0002 Icon 5 Maintenance Guide Version 02 Release date 2020 04 27 Reinstall the metal shield...

Page 10: ...aperture A worn silicone sealing ring will allow the aperture to leak and will form contamination around the module and can even cause errors during measurement due to air or liquid leaks Leaks on ape...

Page 11: ...alves by removing the two screws The ruby is inside the silicone seal Carefully remove it from the aperture half using a tweezer Use a damp cloth immersed into DI water to clean the sides of the apert...

Page 12: ...by is not symmetrical one side is flat the other side has a spherical dent The spherical side must face the LEFT side aperture housing See picture above There is a cut out in the plastic housing where...

Page 13: ...ghtening may cause the silicon to flat out fast again and start leaking The aperture halves must have a small even gap between them Do not flatten out the gap between the aperture halves Doing so will...

Page 14: ...be able to do so use a marker to mark the brake s actual position on two places By doing so it will be easy to find the proper positioning during reassembling Loosen the two small hex screws of the b...

Page 15: ...cap and can get clogged Use the Norma UNF tool not included to manipulate the sampling needle To access the piercing needle the closed wash head must be pushed away from the needle Use a 2mm hex key t...

Page 16: ...there is any dirt residuals are on it Use a brush to clean the wash head in and out Use the white grease from the KIT to lubricate the axis the wash head is turning It is possible to put a little drop...

Page 17: ...tool if needed Overtightening may cause damage to the threads but loose connection will cause leakage in the fluidic system which will show up in erroneous results If there are visible white salt depo...

Page 18: ...state of all injectors look for salt deposits around their bottom fixing plate In case there is any it is maybe a sign of a slight leakage of that injector or of the related valves Check the state of...

Page 19: ...ep 11 Reagent sensors and connections Frequency 1 year or 10000 samples part N A Make sure all tubes are properly placed in the reagent sensors It should click in In case there is a liquid detection p...

Page 20: ...metal tools when replacing the battery you may cause short circuit and a damage the board or the battery Do not touch the surface of the new battery by fingers because contamination salt or moist may...

Page 21: ...following steps Norma iHClean or similar solution is needed for this process Take the Norma iHClean or hypochlorite solution and immerse the diluent pickup tube into it Run diluent prime 3 times and g...

Page 22: ...ion if necessary Run the freshly opened normal control sample and check the NEU XY parameter on the results technical screen It should meet with the given NNEU XY values of the control assay sheet In...