Testing in Timed Mode
1. Refer to “Testing in Steady Mode” on page 17 to purge your dispensing tip with fluid.
2. Change to TIMED mode.
3. Set the time to 000.0250 seconds.
4. Rest the dispense tip on the Dot Standards sheet.
5. Press the foot pedal (or finger switch) to activate the dispense cycle.
NOTE:
The pedal only needs to be pressed for a moment. The complete time will run once you activate the
dispense cycle.
6.
Changing the time changes the deposit size.
If your dot is too small, increase the dispense time to increase
the deposit size. If your dot is too large, decrease the dispense time to decrease the deposit size.
Testing in Teach Mode
1. Refer to “Testing in Steady Mode” on page 17 to purge the dispensing tip with fluid.
2. Change to TEACH mode by going to the menu and scrolling to TEACH.
3. Rest the dispense tip on the Dot Standards sheet.
4. Press the foot pedal to establish your dot size. Depressing the foot pedal repeatedly will have a cumulative
effect on the dot size and dispense time.
5. Continue to press the foot pedal until you achieve the correct deposit size.
6. If needed, press CLEAR to start over.
7. Save the dispense time setting you just created by pressing SAVE. This automatically exits TEACH mode and
returns you to TIMED mode.
Initial Test Procedures (continued)
Ultimus I–II High Precision Fluid Dispensers
18