5
OPERATING INSTRUCTIONS
NOTE:
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
1. Mounting Positions
Securely mount the Pro-Test transducer in the plane of
operation required with the 2 fixing bolts supplied. This can
be to a bench top, wall, or the mounting bracket part number
62198 (not included).
IMPORTANT:
THE WHOLE OF THE TRANSDUCER
MUST ALWAYS BE IN CONTACT WITH
THE SUPPORTING SURFACE.
Clockwise
Only
Calibrating / Testing
Torque tighten to 20
N∙m. (or hand tighten bolts with hex key provided)
Anti-Clockwise
Only
Calibrating / Testing
Torque tighten to 20
N∙m. (or hand tighten bolts with hex key provided)
Clockwise and Anti-Clockwise Calibrating / Testing
IMPORTANT:
BOLTS MUST BE TORQUE TIGHTENED TO 50
N∙m.
REPLACE BOLTS (PART NUMBER 25492) IF REMOVED.
Horizontal
Testing