norbar Pro-Test 1500 SERIES 2 Operator'S Manual Download Page 14

 

 

12 

SERIAL PORT 

 
The serial port is for sending data to a PC or serial printer. 
 

Measurement Mode 

Data Output Operation 

Track 

When ‘PRINT’ is pressed. 

Dial & Electronic 

When ‘ZERO’ is pressed after a peak has been captured. 

Click & Cam 

Automatically after a first peak has been captured. 

 
The data can include direction indication, limit status, measured value, units of measurement and line feed. 
 

Parameters 

 

Parameter 

Options 

Factory Defaults 

Comments 

Parity 

ODD, EVEN or OFF. 

‘OFF’ 

The Parity Bit used for Parity Error. 

Data-Stop bits 

8-2, 8-1, 7-2, 7-1. 

‘8-2’ 

The format per character. 

Baud rate 

1200, 2400, 4800, 9600 or 19200. 

‘9600’ 

The speed of data output. 

Output Limits 

YES or NO. 

‘YES’ 

Limit status sent before data. 

Output line feed 

YES or NO. 

‘NO’ 

Line feed sent after data. 

First character 

- /+ & - /NONE. 

‘-‘ 

Character sent before the data. 

Output units 

YES or NO. 

‘YES’ 

Torque units sent after data. 

 

SET TO FACTORY DEFAULTS. 

 

Reset all settings. 

 
Maximum number of characters per line = 24. 
 
Transmitted data voltage levels are b5 to +9 volts and 

–5 to -9 volts. 

 
Configured as DTE (Data Terminal Equipment) and conforms to RS-232-C specifications. 
 

TIP: 

If the serial port is not communicating with other equipment: 

a)  Check that control word on the Pro-Test and the equipment receiving data match. 

b)  Check that the baud rate is set to the same as the equipment receiving data. 

c)  Check that the connecting lead is wired correctly at both ends if not using the lead 

supplied. 

d)  Check if the equipment receiving data requires the units of measurement inhibited or a 

leading character. 

 
TIP: 

If the serial output is being overwritten set ‘OUTPUT LINE FEED’ to YES. 

 
Hyper Terminal

 

 
The standard Hyper Terminal

 program found in Microsoft

®

 Windows allows the user to view and store serial 

port output data. 

 

 

 

Summary of Contents for Pro-Test 1500 SERIES 2

Page 1: ...norbar com au NORBAR TORQUE TOOLS INC 36400 Biltmore Place Willoughby Ohio 44094 USA Tel 1 866 667 2279 Email inquiry norbar us NORBAR TORQUE TOOLS PTE LTD 194 Pandan Loop 07 20 Pantech Business Hub...

Page 2: ......

Page 3: ...Measurement Mode 7 Fitting Hex Drive Adapter 7 Operating Tool to be Calibrated Tested 7 Set Up 8 Language 8 Limits 8 Units 9 Serial Port 9 Thresholds 9 Modes 9 Limits 10 ISO 6789 1 2017 10 NON ISO 678...

Page 4: ...1500 series 2 TRANSDUCER DISPLAY Assembly Diagram Part Number Description 29034 sq to 10 mm hex socket 29067 sq to 10 mm hex socket 29120 sq to 10 mm hex socket 29079 sq to 22 mm hex socket 29134 sq...

Page 5: ...m hex socket 29079 sq to 22 mm hex socket 29134 sq to 22 mm hex socket 29189 sq to 36 mm hex socket 29179 Operators Manual 34299 Calibration certificate Quick reference card s 34302 Serial data lead 3...

Page 6: ...tolerances The pictorial keys allow direct access to measurement modes Serial Port Connector for data output to PC s or printers PRINT NO PRINT PRINT X PRINT feature for control of serial port output...

Page 7: ...the mounting bracket part number 62198 not included IMPORTANT THE WHOLE OF THE TRANSDUCER MUST ALWAYS BE IN CONTACT WITH THE SUPPORTING SURFACE Clockwise Only Calibrating Testing Torque tighten to 20...

Page 8: ...4 Switching On Switch ON Pro Test I in the down position WARNING ALLOW PRO TEST TO EQUALISE TO THE AMBIENT TEMPERATURE HUMIDITY BEFORE SWITCHING ON WIPE OFF ANY MOISTURE BEFORE USE TIP If there is no...

Page 9: ...TRONIC Bending beam wrenches Dial type wrenches Hold highest signal Reset by pressing CLICK CAM Preset click type wrenches Captures the FIRST PEAK of torque and displays this for the selected AUTO RES...

Page 10: ...for next menu Press EXIT to end set up LANGUAGE OK EXIT ENGLISH DANSK FRANCAIS NEDERLANDS DEUTSCH SUOMI ITALIANO NORSK ESPANOL SVENSKA PORTUGUES LIMITS OK EXIT X ISO 6789 1 2017 X NON ISO 6789 1 2017...

Page 11: ...lb gf cm lbf ft in oz lbf in Kgf m X MODES OK EXIT DIAL ELECTRONIC CLICK CAM X SERIAL PORT OK EXIT PARITY OFF DATA STOP BITS 8 2 BAUD RATE 9600 OUTPUT LIMITS YES EVEN ODD OFF SERIAL PORT OK EXIT OUTPU...

Page 12: ...t value x 100 Target value If the 100 target value is set above 10 N m the deviation must be within 4 For a 100 target value of 10 N m or below the deviation must be within 6 Example For a 100 N m tar...

Page 13: ...ARGET is set to 4 See the SET UP section to change the default value Operation Measurement Mode Limit Operation Track Limits follow signal and are not held Dial Electronic Limits status is held until...

Page 14: ...haracter sent before the data Output units YES or NO YES Torque units sent after data SET TO FACTORY DEFAULTS Reset all settings Maximum number of characters per line 24 Transmitted data voltage level...

Page 15: ...m LO 1 0335 N m NONE Clockwise 1 0335 N m LO 1 0335 N m Anti Clockwise 1 0335 N m LO 1 0335 N m Pin Connections Pin No Function 1 Not Connected 2 Received data to Pro Test 3 Transmitted data from Pro...

Page 16: ...is functioning at maximum accuracy are available Cleaning Do not use abrasives or solvent based cleaners Disposal Recycling Considerations Component Material Back Panel Stainless Steel Transducer Sha...

Page 17: ...ture range 5 C to 40 C Storage temperature range 20 C to 70 C Maximum operating humidity 85 Relative Humidity at 30 C Power supply 90 to 264 Volts a c at 50 60 Hz input 9V 300 mA D C output centre pos...

Page 18: ...16 NOTES...

Reviews: