24
Figure 13: Squaring the Ball Target to the Mounting Ring
11.6
Place the square on the sensor ring as shown in
Figure 14
and follow
the procedure in Step 11.5.
Figure 14: Squaring the Ball Target to the Mounting Ring
11.7
Repeat the process until the ball target is centered 20.5” from the
sensor mounting ring.
1/16”
(1.5mm)