Precautions for Use 

Warnings 

• 

The Model 8500AV Pulse Oximeter is intended for VETERINARY USE ONLY.  Do not use on human patients. 

• 

Accurate oxygen saturation measurement cannot be obtained when the oximeter is not measuring the pulse properly.  
If the Perfusion LED is erratic or the Pulse Rate display is erratic or inaccurate, first examine the animal for any 
signs of distress and only then reexamine sensor placement. 

• 

Do not operate in an explosive atmosphere. 

• 

Inadequate perfusion, thick fur, foreign matter that blocks light, or an improperly applied sensor can result in erratic 
and inaccurate oxygen saturation and/or pulse rate measurement.  Follow the instructions outlined in “Verify 
Operation.”  If proper operation cannot be verified, remove the sensor from the animal and DO NOT use the 
oximeter on this animal. 

• 

In the event the sensor becomes dislodged from the animal, audible and visual alarms are activated, requiring that a 
veterinary professional investigate the reason for the alarm status. The veterinary  professional must investigate 
animal status and sensor attachment after every sensor alarm indication. It is possible when the sensor is dislodged 
from the animal (under certain conditions of light and vibration of the sensor) for the pulse oximeter to display 
normal physiological values.  

• 

NONIN does not recommend operating the Model 8500AV in an MRI environment. 

• 

Do not remove any covers other than the battery cover when battery replacement is necessary.  There are no user 
serviceable parts inside other than the replaceable batteries.  Refer service to qualified NONIN personnel. 

• 

Operation of this device below the minimum amplitude of 0.5% modulation may cause inaccurate results.  

• 

This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, 
the device should be observed carefully to verify normal operation. 

• 

The use of accessories, sensors, and cables other than those listed in this manual may result in increased emission 
and/or decreased immunity of this device. 

Cautions 

• 

Read this manual carefully before using the Model 8500AV Pulse Oximeter. 

• 

USE ONLY NONIN VETERINARY SENSORS. 

• 

Use only NONIN compatible accessories and veterinary sensors to preserve the integrity and accuracy of the 
monitor. 

• 

Cardiogreen and other intravascular dyes, depending on concentration, could affect the accuracy of the 
measurement. 

• 

The Model 8500AV Pulse Oximeter is calibrated to determine the percentage of arterial oxygen saturation of 
functional hemoglobin measured at the finger tip of an adult human. Significant levels of dysfunctional 
hemoglobins such as carboxyhemoglobin or methemoglobin might affect the accuracy of the measurement. Other 
types of hemoglobin and alternate sensor locations may affect the calibration. 

• 

Certain pharmacologic agents used to sedate or anesthetize animals may have cardiovascular effects that can 
adversely affect the performance of the pulse oximeter by reducing the perfusion to the sensor site. Examples of 
commonly used agents that may have this type of effect on certain animal species are Detomidine HCl and 
Xylazine HCl. 

• 

There is a wide range of variability between animal species and their respective differences in anatomy, physiology 
and responses to veterinary pharmacological agents. Therefore, the veterinarian will need to use discretion when 
selecting sensors and/or sensor sites that are appropriate for the animal species and the monitoring conditions. 

• 

When attaching the sensor, make sure not to secure the sensor in a manner that will restrict perfusion. This could 
inhibit proper function of the oximeter and cause discomfort or localized ischemia to the animal.   

• 

If the sensor is not positioned on a perfused site, or at least ten seconds of green perfusion pulses cannot be 
observed for a given sensor, change sensor location or sensor type until this condition is achieved. 

Summary of Contents for 8500AV

Page 1: ...Operator s Manual Model 8500AV Alarm Hand Held Veterinary Pulse Oximeter Nonin Medical Inc 13700 1st Avenue North Plymouth MN 55441 5443 763 553 9968 1 800 356 8874...

Page 2: ...recautions for Use 1 Introduction 3 Operating Instructions 4 General 4 Connect Sensors 4 Turn on the Oximeter 4 Verify Operation 4 On Alarm Models Only 4 Veterinary Sensors 5 Sensor Placement 6 Batter...

Page 3: ...carefully to verify normal operation The use of accessories sensors and cables other than those listed in this manual may result in increased emission and or decreased immunity of this device Caution...

Page 4: ...all equipment must be installed and put into service according to the EMC information specified in this manual This device has not been tested for immunity to electromagnetic disturbances Portable and...

Page 5: ...determine arterial saturation and pulse rate The Model 8500AV requires no routine calibration or maintenance Oxygen saturation and heart rate are displayed on light emitting diode LED digital display...

Page 6: ...the perfusion light If the perfusion light is red or yellow reposition the sensor or try a different sensor If normal operation cannot be achieved call NONIN Customer Support for assistance After eigh...

Page 7: ...nitoring These sensors are manufactured to meet the calibration requirements for the NONIN Model 8500AV Pulse Oximeter The oximeter is calibrated for adult human hemoglobin measured at the finger tip...

Page 8: ...easurement error Secure the sensor with tape ensuring that the tape does not restrict perfusion Model 2000T Transflectance Sensor Application The 2000T Transflectance sensor is designed for placement...

Page 9: ...the left hand display indicating the Alarm mode dft will appear in the right hand display Each time the arrow switch is pressed the option on the right hand display will advance When the correct valu...

Page 10: ...h nn 00 through 12 Day d 01 through 31 Hours h 00 through 23 Minutes nn 00 through 59 Printer mode is used by the model 8500P Printer WARNING The use of the 8500P may result in increased emission and...

Page 11: ...hat may affect the SpO2 data The following is a simple illustration of how to interpret the perfusion LED Perfusion LED blinks GREEN pulse waveform signal is of good quality and SpO2 data is accurate...

Page 12: ...splayed or when the unit is out of track Press the 1 switch while holding the arrow switch This will cause the current settings of the four alarms to be displayed in the same format and order as durin...

Page 13: ...Cathode 3 Infrared Cathode Red Anode 4 Serial Data TTL Levels 5 Detector Anode 6 Logic Level 7 Cable Shield 8 Coax Shield 9 Detector Cathode 5V a Serial format Real time mode ASCII 9600 BAUD 9 Data B...

Page 14: ...cale for adults using or finger clip sensors Below 70 not specified b Rate Accuracy 3 1 digit 7 Temperature a Operating 20 to 50 C b Non Operating 30 to 50 C 8 Humidity a Operating 10 to 90 Non Conden...

Page 15: ...e work should be done under a microscope set at 25 times magnification Use of regular soldering equipment will damage the board and void any applicable warranty Nonin will repair out of warranty oxime...

Page 16: ...systems or accessories delivered to the purchaser which are found to be defective in any manner whether such remedies be in contract tort or by law This warranty excludes cost of delivery to and from...

Page 17: ...er Model 8500MB Mounting Bracket Wall or Pole mount Model 8500CC Carrying Case Black or Yellow Model 8500I Cable 3 foot extension Model 8500VI Veterinary Cable 10 foot extension Model 8500PCC Printer...

Reviews: