Operator’s Manual 

 

Model 8500AV (Alarm) 

Hand-Held Veterinary Pulse Oximeter 

 

 

 

 
 

 
 
 

 
 
 

 

Nonin Medical, Inc. 

 13700 

1

st

 Avenue North 

 

Plymouth, MN 55441-5443 

 (763) 

553-9968 

 1-800-356-8874

 

 
 

Summary of Contents for 8500AV

Page 1: ...Operator s Manual Model 8500AV Alarm Hand Held Veterinary Pulse Oximeter Nonin Medical Inc 13700 1st Avenue North Plymouth MN 55441 5443 763 553 9968 1 800 356 8874...

Page 2: ...recautions for Use 1 Introduction 3 Operating Instructions 4 General 4 Connect Sensors 4 Turn on the Oximeter 4 Verify Operation 4 On Alarm Models Only 4 Veterinary Sensors 5 Sensor Placement 6 Batter...

Page 3: ...carefully to verify normal operation The use of accessories sensors and cables other than those listed in this manual may result in increased emission and or decreased immunity of this device Caution...

Page 4: ...all equipment must be installed and put into service according to the EMC information specified in this manual This device has not been tested for immunity to electromagnetic disturbances Portable and...

Page 5: ...determine arterial saturation and pulse rate The Model 8500AV requires no routine calibration or maintenance Oxygen saturation and heart rate are displayed on light emitting diode LED digital display...

Page 6: ...the perfusion light If the perfusion light is red or yellow reposition the sensor or try a different sensor If normal operation cannot be achieved call NONIN Customer Support for assistance After eigh...

Page 7: ...nitoring These sensors are manufactured to meet the calibration requirements for the NONIN Model 8500AV Pulse Oximeter The oximeter is calibrated for adult human hemoglobin measured at the finger tip...

Page 8: ...easurement error Secure the sensor with tape ensuring that the tape does not restrict perfusion Model 2000T Transflectance Sensor Application The 2000T Transflectance sensor is designed for placement...

Page 9: ...the left hand display indicating the Alarm mode dft will appear in the right hand display Each time the arrow switch is pressed the option on the right hand display will advance When the correct valu...

Page 10: ...h nn 00 through 12 Day d 01 through 31 Hours h 00 through 23 Minutes nn 00 through 59 Printer mode is used by the model 8500P Printer WARNING The use of the 8500P may result in increased emission and...

Page 11: ...hat may affect the SpO2 data The following is a simple illustration of how to interpret the perfusion LED Perfusion LED blinks GREEN pulse waveform signal is of good quality and SpO2 data is accurate...

Page 12: ...splayed or when the unit is out of track Press the 1 switch while holding the arrow switch This will cause the current settings of the four alarms to be displayed in the same format and order as durin...

Page 13: ...Cathode 3 Infrared Cathode Red Anode 4 Serial Data TTL Levels 5 Detector Anode 6 Logic Level 7 Cable Shield 8 Coax Shield 9 Detector Cathode 5V a Serial format Real time mode ASCII 9600 BAUD 9 Data B...

Page 14: ...cale for adults using or finger clip sensors Below 70 not specified b Rate Accuracy 3 1 digit 7 Temperature a Operating 20 to 50 C b Non Operating 30 to 50 C 8 Humidity a Operating 10 to 90 Non Conden...

Page 15: ...e work should be done under a microscope set at 25 times magnification Use of regular soldering equipment will damage the board and void any applicable warranty Nonin will repair out of warranty oxime...

Page 16: ...systems or accessories delivered to the purchaser which are found to be defective in any manner whether such remedies be in contract tort or by law This warranty excludes cost of delivery to and from...

Page 17: ...er Model 8500MB Mounting Bracket Wall or Pole mount Model 8500CC Carrying Case Black or Yellow Model 8500I Cable 3 foot extension Model 8500VI Veterinary Cable 10 foot extension Model 8500PCC Printer...

Reviews: