41
Technical Information
LOW PERFUSION
Reusable:
Finger Clip:
±3 digits
±3 digits
Flex:
±3 digits
±3 digits
Soft Sensor:
±3 digits
±3 digits
Reflectance:
±3 digits
±3 digits
Disposable:
6000 Series:
±3 digits
±3 digits
7000 Series:
±3 digits
±3 digits
Measurement Wavelengths and Output Power:
b
Red: 660 nm @ 0.8 mW maximum average power
Infrared: 910 nm @ 1.2 mW maximum average power
Memory
70 hours (assuming continuous operation)
Temperature (Operating):
0º to +40º C (32º F to 104º F)
Temperature (Storage/Transportation):
-30º to +50º C (-22º F to 122º F)
Humidity (Operating):
10 to 90% noncondensing
Humidity (Storage/Transportation):
10 to 95% noncondensing
Altitude (Operating):
Up to 12,000 meters (40,000 feet)
Hyperbaric Pressure
Up to 4 atmospheres
Power Requirements (Mains)
100-240 VAC 50-60 Hz
Internal Power
Battery: 7.2 volt NiMH battery pack
Operating Life (fully charged
battery):
16 hours minimum
Storage Life: 21 days minimum
Recharge Rate: 4 hours maximum
Dimensions
Approximately 219 mm (8.6”) W x 92 mm (3.6”)
H x 142 mm (5.6”) D
Weight
Approximately 900 grams (2 lbs) with battery
Warranty
3 years
Classification per IEC 60601-1/CSA601.1/UL60601-1:
Type of Protection: Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection:
IPX2
Analog Outputs:
SpO
2
Output Range: 0-1 VDC (0-100% SpO
2,
1.27 VDC (out of track)
Pulse Rate Output Range: 0-1 VDC (0-300 BPM),
1.27 VDC (out of track)
Event Maker: 0 V (no event), 1 V (event occurred)
Accuracy: ±2% (SpO
2
), ±5% (Pulse Rate)
Load Current: 2 mA maximum
b.) This information is especially useful for clinicians performing photodynamic therapy.