15
Operator Functions
Table 1: Limits Display Sequence
Parameter
Parameter
(SpO
2
)
Display
Initial
Setting
(Pulse Rate
Display)
Adjustment Range
Recall Alarm Settings
“
rCL
”
“
no
”
“
yES
” or “
no
”
Low %SpO
2
Alarm Limit
“
02L
”
2,3
“
85
” “
Off
”, 50 to 95 by 1
Pulse High Alarm Limit
“
HH
”
2
“
200
” “
Off
”, 75 to 275 by 5
Pulse Low Alarm Limit
“
HL
”
2
“
50
” “
Off
”, 30 to 110 by 5
High %SpO
2
Alarm Limit
“
02H
”
2
“
Off
” “
Off
”, 80 to 100 by 1
Alarm Volume
“
adb
”
2
“
Hi
” “
Off
” or “
Lo
” or “
Hi
”
Clear Memory
“
Clr
”
1
“
no
” “
yES
” or “
no
”
Confirm Memory Clear
“
del
”
1
“
no
” “
yES
” or “
no
”
Year
“
y
” “
00
”
0 to 99 by 1
Month
“
nn
” “
00
”
0 to 12 by 1
Day
“
d
” “
00
”
1 to 31 by 1
Hour
“
h
” “
00
”
0 to 23 by 1
Minutes
“
nn
” “
00
”
0 to 59 by 1
Notes:
1) Both of these menu options are part of the memory clear command; “
del
” will be
displayed only if “
yES
” was selected as the setting for the “
Clr
” parameter.
2) These parameters are restored when Recall Alarm Settings is set to “
yES
.” these are
also the settings displayed by Review Alarm Settings.
3) The low SpO
2
Alarm limit saved for recall cannot be lower than the current default for that
alarm limit. If it is, the default value will be used when alarm limits are restored.