CHECKING THE ACCURACY

1.

 Place the camera at a distance of 1.0 m (distance from the front side of the NL Sonic Tester to the front 

side of the NL Camera). Make sure that the NL Camera is aligned parallel to the NL Sonic Tester.

2. 

Align the NL Camera in such a way that the calibrator is exactly within the outline in the middle of the 
screen and the IR LED (

2

) is within the circle and it is illuminated at its brightest. This means that the 

calibrator and camera are aligned properly.

3

. Observe the 

dB indicator bar

 on the right side of the screen. The NL Camera accuracy is correct if the reading 

is within the green range. If the reading is in the red range, first make sure that the NL Camera and NL Sonic 

Tester are set up correctly and that the IR LED is at its brightest. If the device still displays incorrect (red) 
readings, please contact your distributor.

Summary of Contents for NL SONIC TESTER SC10

Page 1: ...SONICTESTERSC10 User Manual...

Page 2: ...Noiseless Acoustics Oy Sitratie 7 00420 Helsinki Finland 358 0 10 583 3240 info nlacoustics com www nlacoustics com FI26650896...

Page 3: ...e 5 Disposal of Electronic Waste 5 Markings 5 Mechanical Environmental 5 Device Parts 6 Getting Started 7 Tripod 7 Startup 7 Shutdown 7 Operating Modes 7 Using the Device for Checking the NL Camera Ac...

Page 4: ...hould always use hearing protection when the device is operated at a distance less than 0 5 m to avoid permanent hearing damage NOTE Sound pressure level may exceed 80 dB A at a distance less than 0 5...

Page 5: ...ment marked with the crossed out wheeled bin above is electrical and electronic equipment The crossed out wheeled bin symbol indicates that waste electrical and electronic equipment should not be disc...

Page 6: ...6 Device Parts FRONT SIDE BACK SIDE 1 Loudspeaker 2 IR LED 3 Mode switch 4 QR plate 5 Power LED 6 Power socket 1 2 3 5 6 4...

Page 7: ...into a wall socket before connecting the DC barrel connector to the power socket 6 of the device After connecting the DC barrel connector the device will take about 10 seconds to start NOTE If the ma...

Page 8: ...on your specific camera configuration If you accidentally miss the page containing the selection keep tapping until it appears again 5 Power on the NL Sonic Tester 6 Mount the device on a tripod Place...

Page 9: ...f the screen and the IR LED 2 is within the circle and it is illuminated at its brightest This means that the calibrator and camera are aligned properly 3 Observe the dB indicator bar on the right sid...