
- 13 –
SL50
User guide – 33-60-022-000
2 CALIBRATION OF SENSORS
Every
nke
sensor is adjusted at the factory. However, a calibration is required to adapt the
sensor to the specificities of your boat and to obtain an optimum measurement accuracy.
Follow the calibration procedure below, by visualising the settings on a display.
Please refer to the installation notice of the
TOPLINE
sensor that you wish to calibrate.
2.1 Setting procedure of calibration coefficient
Example : you wish to set up the calibration coefficient of the
boat speed
channel at this value :1.1.
2.2 Offset setting procedure
Follow the above procedure and choose the
OFFSET
sub-channel.
Note that this value is set to
0.
1
2
Using the keys
3
4
Figure 4
Press
The calibration is saved in memory.
Ent
Then successively press to select the sub-channel
CALIB COEF.
nke
SL50
kts
UNIT
nke
SL50
1.00
CALIB COEF
nke
SL50
16.85
BOAT SPEED
select the cannal from which you wish to select the calibration coeficient.
until the list of sub-channels
UNIT, CALIB COEF
and
OFFSET
appears.
Ent
nke
SL50
16.85
BOAT SPEED
Ent
Press to validate and exit by maintaining pressure on . .
Ent
nke
SL50
1.00
CALIB COEF
Using the keys
change the new calibration coefficient to
"1.1"
.
5
CAUTION :
-
The
calib coef
parameter is a multiplier coefficient. This value must never be equal to
zero. By default this coefficient is set to
1.00
. If it is not the case, before starting a
calibration enter the value
1.00.