Instruction manual
OCM Pro CF
®
OCM Pro CF - Rev. 09 as of 07.06.2016
page 139
Fig. 9-131 Indication of measured single velocities
+
Toggle between both pages (measurement windows 1-8 and 9-16) by using the
up and down keys.
Depending on the sensor type used the first measuring window can be found at
varying height level positions.
A reading of ------ in a measurement window indicates that there is currently no
flow velocity able to be measured in the according window. This might happen
due to very clean water or vorticity within this area. This effect might occur as
well in case of low flow levels, however is caused due to the OCM Pro automati-
cally reducing the number of measurement windows here.
The measurement result will not be influenced even if one or few windows
should fail!
H sensors
This menu indicates the measured filling levels. Depending on the used sensors
for level measurement (via water-ultrasonic, pressure, air-ultrasonic or 2-wire
sensor, see chapter 9.5.2) different menus are displayed:
Example 1:
1 Level
2 Level water-ultrasound
3 Level pressure internal
4 Level air-ultrasound NIVUS
Fig. 9-132 Menu selection with water-ultrasonic, pressure and
air-ultrasonic