Instruction manual
NivuSonic Clamp on
®
A correct and reliable measurement can be performed only in full filled pipes.
For this reason do not install measurements in downpipes or at the highest point
of the pipeline.
1 = recommended range in almost horizontal position
(sensors can be installed on the side)
2 = recommended range in vertical pipe
3 = not recommended due to part filling/idling
4 = measuring impossible due to idling
Fig. 6-10
Comparison of installation places
In case of planning measurements in horizontal pipelines we recommend to
consider a slightly inclined section or an inverted siphon.
or
Fig. 6-11
Horizontal pipe with inverted siphon
NivuSonic CO - Rev. 00 as of 08.12.2010
Page 27