Universal inputs
(PWM mode 1/2)
PWM
(1/2)
Filter time
Minimum signal length for signal detection
Decay
Temporal function in the measurement cycle
Off
Decay deactivated
Min
The minimum of the last x measurement
values is recorded.
Max
The maximum of the last x measurement
values is recorded.
Avg
The arithmetic mean of the last x
measurement values is recorded.
Med
The median of the last x measurement
values is recorded.
Rms
The root mean square of the last x
measurement values is recorded.
Time
Time window for the decay. Up to max. 60 measurement
values can be used for the calculation (e.g. 1 channel: 60
measurement values; 4 channels: per 15 measurement
values). To calculate the number of considered
measurement values see "Example for explaining the
recording, measuring and burst cycles (in conjunction with
the decay)" on page 59.
Hold
Maintain the last valid measurement value for x
measurement cycles
Off
Function deactivated
1-5
Number of measurement cycles for which
the measured value is maintained until the
error value is issued
On
In the event of an error, the last valid
measurement value is maintained until a new
valid measurement value is present.