Thermo Scientific AquaSensors
™
AquaClear
™
Low Range Turbidimeter User Guide
23
When the temperature calibration procedure is started, the analog output is placed into hold mode. The
user is prompted to prepare the sensor. The sensor value is dynamically updated during this step.
When the sensor is ready, the ENTER key is pressed and the user is presented with a calibration value for
editing. Please note that it is best to calibrate temperature when the sample chamber is full of flowing process
water.
When the value has been edited as desired, the enter key is pressed and the calibration of the point is started.
During this time, the calibration procedure can be aborted by pressing the escape key.
After the calibration point has been stored, the user is prompted to confirm a successful calibration procedure.
Temperature
Calibrating...
MENU
ESC
ENTER
Temperature?
( 25.0) °C
MENU
ESC
ENTER
Temperature?
( 25.3) °C
MENU
ESC
ENTER
25.3 °C
Sensor ready?
MENU
ESC
ENTER