Thermo Scientific AquaSensors
™
AquaClear
™
Low Range Turbidimeter User Guide
21
From the Calibrate menu, use the UP/DOWN arrows to display ZERO calibration. Press ENTER to select.
Be sure the filtered water has had an opportunity to flow for a few minutes to clean out any particles that may
have been in the plumbing. Press ENTER to initiate zero calibration.
A zero calibration may take some time, as the zero must be calibrated over a wide dynamic range. During this
process the calibrating… message is shown.
When the zero calibration is complete the user will be asked to confirm the result with the ENTER key.
Zero
Confirm cal ok?
MENU
ESC
ENTER
Zero
Calibrating…
MENU
ESC
ENTER
0.020 NTU
Sample Ready?
MENU
ESC
ENTER
Calibrate
Zero
MENU
ESC
ENTER
1-Point Sample
Temperature
Exit