5
18. Again select NO scrubber
19. Input the passcode 0000
using the number buttons
to confirm that a new
NO scrubber is installed
20. Select OK to accept the
changes
21. Optional: Activate
Bluetooth for wireless
communication with PC or
connect a USB cable, see
page 18
22. Select Return to go back
to the Main screen
23. Allow the instrument to
stabilize.
Note!
Stabilization of the instrument normally may
take up to 30 minutes following connection
of the power supply unit to the power outlet.
However, it is recommended to allow a new
NIOX MINO
®
Sensor to be connected for two
hours before the first measurement is done.
Measurement
Depending on what is shown on the display,
proceed as follows:
Stabilization in progress
Wait until ready
Instrument in sleep mode
Touch the display
Ready for use
Enter patient ID (optional)
Note!
If Patient ID is used, it has to be entered
before each measurement (even if it is the
same patient).
Input a patient specific ID number, up to 10 digits.
1. Select the ID button on
the main screen
2. Input the patient specific
ID using the number
buttons
to
Use the shift buttons to
change between digits 0
to 4 and 5 to 9
3. Select OK to accept the
patient ID