11
8. Input the passcode 0000
using the number buttons
to confirm that a new
NO scrubber is installed
9. Select OK
Disposal of used/expired products
• Used or expired NIOX MINO
®
Sensors should be
recycled according to the local recycling program
for electronic equipment. Note that there is a
LiMnO
2
battery inside the Sensor unit*.
• An expired NIOX MINO should be recycled,
according to the local program for electronic
equipment. Note that there is a LiMnO
2
backup
battery inside the instrument*.
• The NO scrubber contains potassium
permanganate and should be disposed of as
hazardous waste in accordance with the local
waste disposal regulations.
NIOX MINO is RoHS compliant.
* The batteries are not user-replaceable parts.
Return shipments
Contact your local Circassia representative or
Circassia AB.
Support
Please contact your local Circassia representative
or Circassia Technical Support if you encounter
problems, which you can not solve with the actions
stated in this manual.
For contact details, see back-page, and provide the
following information:
• Your name, address and telephone number.
• Serial No. (both instrument and Sensor).
• Problem description (as thorough as possible).
• Alert codes or lists.
Limited warranty
Circassia AB provides a Limited Warranty for this
instrument and original accessories delivered with
this instrument. Conditions are defined when the
items are purchased.
Do NOT try to repair the instrument. It is NOT
permitted to open the instrument. Any attempt
will make the warranty invalid and performance
according to the specifications can not be
guaranteed.
Troubleshooting
Warning
Action
The inhalation was too weak
to initiate a measurement
or an exhalation into the
instrument was performed
prior to an inhalation - Stop
the procedure immediately
when this warning appears.
Wait until the main menu
screen is displayed and repeat
the inhalation with a stronger
inhalation force.
NO scrubber reminder.
The symbol is shown at first
start-up of the instrument
as a reminder to insert and
set the software for a new
NO scrubber. See
Installation
and set-up
section page 4.
NO scrubber almost expired -
Order a new NO scrubber.
The symbol is shown when
10% of the measurements are
remaining or 2 weeks before
expiration date and continue
until the NO scrubber has
expired. A NO scrubber can be
used for 1000 measurements
or 1 year. Refer to the Change
NO scrubber section page 10.
No Sensor connected -
Connect a Sensor.
Sensor almost expired – Order
a new Sensor.
The symbol is shown when
10% of the measurements
are left or 2 weeks before
expiration date and continue
until the Sensor has expired.
Refer to the Change Sensor
section page 10.