EU Patient Manual --560-IFU-IT1098, Rev. A 25
Max Vol
Reached,
Pump
Standby
The pump has
reached a
maximum volume
per hour or
interval limit.
Note
:
The
infusion will be
stopped
automatically.
The screen
displays “Max Vol
Reached, Pump
Standby
” until the
new hour or
interval is reached
when the infusion
can begin again;
the LED shows a
solid yellow light.
The pump will remain in
standby mode until the new
hour or interval is reached
before beginning the
infusion. No user action is
required.
To change a volume limit,
hold the RUN/STOP key for
3 seconds to stop the
infusion. Scroll down to
highlight the Max Vol/h or
Max Vol/Int parameter.
Press the ENTER key to
change the parameter.
Enter the Clinician Code to
unlock the prescription.
Change the value. Press
ENTER to confirm the new
parameter. Press the
RUN/STOP key to activate
the change in the therapy.
Invalid
Infusion
Parameter
The parameter
value entered
exceeds the
remaining pump
life.
The screen
displays “Invalid
Infusion
Parameter”;
the LED flashes a
red light;
the audio alert
sounds.
Press and release the
RUN/STOP key to
acknowledge the condition
and silence the audio alert.
Enter a valid infusion
parameter value and start
the infusion as applicable.
Infusion
Complete
The infusion is
completed.
The screen
displays “Infusion
Complete”;
the LED flashes a
red light; the audio
alert sounds.
Press the RUN/STOP key
to acknowledge the
condition and silence the
audio alert.
If the alert condition is not resolved by following the above instructions, the user should
stop using the pump immediately and contact the pump provider for further instructions.