________________________________________________________________
Nimbus
TM
EU Patient Manual --560-IFU-IT1098, Rev. A 23
Downstream
Occlusion
Increased
pressure is
sensed in the line
between the
pump and the
patient.
Note
:
The
infusion will be
stopped
automatically.
The screen
displays
“Downstream
Occlusion”; the
LED flashes a red
light; the audio
alert will sound
immediately
(optional delay) if
the downstream
occlusion is not
resolved.
Press and release the
RUN/STOP key to silence
the audio alert (if present).
Resolve the cause of the
occlusion.
When resolved, the pump
will automatically resume
the infusion.
Pump
Unattended
The pump is
standing by and
not infusing for
more than 10
minutes.
The screen
displays “Pump
Unattended”; the
LED shows a solid
yellow light; the
audio alert
sounds.
Press and release the
RUN/STOP key to
acknowledge the condition
and silence the alert. The
pump will return to the
programming screen. Press
the RUN/STOP key then
follow the prompts to start
or resume the infusion. Or
press and hold the ON/OFF
key for 3 seconds to power
off the pump.
Note
: The audible alert will
return after 10 minutes if no
further action is taken after
acknowledging the alert.
System Error
The pumping
mechanism may
not work properly.
Note:
The
infusion will be
stopped
automatically
.
The screen
displays “System
Error”; the LED
flashes a red light;
the audio alert
sounds.
Press and release the
RUN/STOP key to
acknowledge the condition
and silence the alert. Power
off the pump. Take the
pump out of service. Use a
replacement pump if
needed.