5 Assembly
1 Assembling the Basic Set
E-34
2
Replacing fuses (for TS100/TS100-F only)
In order to prevent electrical shock and/or damage to equipment, be sure to turn
off the microscope's power (press the power switch to the “
○
” position) and
unplug the power cord before replacing fuses.
Use only the specified fuses. Use of other fuses
may result in malfunction or fire.
Specified fuse: Two 250-V, 1-A time-lag,
low-breaking 5x20 miniature fuses
(1)
Press the power switch to the “
○
” position
and unplug the power cord if it is connected.
(2)
Remove the fuse holder using a flathead screwdriver. (See P.33.)
(3)
Remove the fuses and replace with new fuses.
(4)
Check that the voltage displayed in the fuse holder window matches the actual input
voltage.
(If it differs, change the voltage selector setting according to instructions on P.33.)
(5)
Re-attach the fuse holder in its original location.
3
Attaching (replacing) a lamp (for TS100/TS100-F only)
In order to prevent burn injury and/or electrical shock when replacing a lamp, be
sure to turn off the power (press the power switch to the “
○
” position), unplug
the power cord, and allow the lamp and its surrounding areas to cool sufficiently
before opening the lamphouse cover.
Use only the specified lamp. Use of another type of
lamp may result in malfunction or fire.
Specified lamp: 6 V-30 W halogen lamp
(PHILIPS 5761)
(1)
Remove the lamphouse cover by lightly
pressing on the front and rear hem and
pulling upward.
(2)
Handling the lamp using gloves or piece of
cloth, firmly plug the lamp into the socket pin
holes as far as it will go.
(3)
Return the lamphouse cover to its original
position.
(4)
Plug the lamp cable connector into the socket
on the rear of the microscope.
Fuses Voltage selector
Voltage display
Window
WARNING
WARNING