64
7. List of Control Commands
Identification
code
c
c
c
c
r
r
r
c
r
s
c
Specifications
Rotates the nosepiece in forward direction to the next address.
Rotates the nosepiece in reverse direction to the next address.
(However, rotation from nosepiece address 1 to 5 is prohibited.)
Rotates the nosepiece in reverse direction to the next address.
Rotates the nosepiece to the specified address (p: 1 to 5).
Reads the nosepiece address.
Reads the program version.
Reads the program name.
Sets the status output setting.
Reads the status output setting.
Outputs the status.
Initializes the control data (factory default).
Command
RCW
RCR
RCC
RDC
RAR
VER
PNM
SAS
SAR
SAE
DEF
Data
-
-
-
p
-
-
-
-
-
-
-
8. Response to Control Commands
[c] [RCW] [<CR>]
Rotates nosepiece in forward direction to the next address.
→
When properly finished [o] [RCW] [<CR>]
→
When control error occurred [n] [RCW] [error code] [<CR>]
[c] [RCR] [<CR>]
Rotates nosepiece in reverse direction to the next address. (Rotation from address 1 to address
5 is prohibited.)
→
When properly finished [o] [RCR] [<CR>]
→
When control error occurred [n] [RCR] [error code] [<CR>]
[c] [RCC] [<CR>]
Rotates nosepiece in reverse direction to the next address.
→
When properly finished [o] [RCC] [<CR>]
→
When control error occurred [n] [RCC] [error code] [<CR>]
[c] [RDC] [p] [<CR>]
Rotates nosepiece to the specified address (p: 1 to 5).
→
When properly finished [o] [RDC] [<CR>]
→
When control error occurred [n] [RDC] [error code] [<CR>]
[r] [RAR] [<CR>]
Reads nosepiece address.
→
When properly finished [a] [RAR] [p] [<CR>]
→
When control error occurred [n] [RAR] [error code] [<CR>]
[p]: Nosepiece address 0, 1, 2, 3, 4, or 5 (“0” when address unidentified.)