4
Assembly
37
IC Inspection Microscope ECLIPSE L200ND / L200N Instructions
Fi
b
er adapter clamp scre
w
Fi
b
er clamp scre
w
Light g
u
ide fi
b
er
Figure-43
X
Connecting the microscope and an external light source
CAUTION
When the microscope is used with an external light source, be sure to attach the C-HGFIE HG
Precentered Fiber Illuminator (motorized type). The shutter must be controlled in synchronization
with the motorized nosepiece. Connect the microscope and the C-HGFIE by plugging the cable to
the HGFIE connector on the rear of the microscope and the RS232C connector on the C-HGFIE.
If this communication is not established, a flash of light may be fired at the rotation of the
motorized nosepiece. Be sure to connect them.
Microscpe main body (rear view)
C-HGFIE (rear view)
Figure-44 Connecting the microscope and the illuminator
HGFIE connector
RS232C
connector
Fiber adapter
RS232C
cross cable
Light
guide fiber