EP1300
EC-METER EC1200 AND PH-METER PH1200
_____________________________________________________________________________________________
9
4.2 TECHNICAL SPECIFICATIONS
The DISP number next to the standard default shows the location of the data in the menu.
SETUP parameters are displayed as: “S x.y”
Default
DISP
1.0 SECONDARY MEASURE TEMPERATURE
EC (conductivity)
1.01
Input EC1200:
2-wire K=1 cel, art.no. EC5005
EC range: Resolution: under value: max. range:
20.00 µS
0.01 µS
-1.00 µS
21.00 µS
200.0 µS
0.1 µS
-10.0 µS
210.00 µS
2000 µS
1 µS
-100 µS
2100 µS
20.00 mS
0.01 mS
-1.00 mS
21.00 mS
200.0 mS
0.1 mS
-10.0 mS
210.0 mS
Under range:
-<<<<uS/ mS
Above range:
>>>>uS/ mS
Zero:
+/- 10 %
0 %
1.1
Calibration of zero is performed for each range starting from the
lower one.
Sens:
60/160%
100%
1.2
Calibration:
manual/automatic with KCL standard liquid
0.01N
0.1N
1N
Standard
Nieuwkoop
4.0 liquid
Temp. ref.
20°C
1278 µS
11.67 mS 102.1 mS
3.6 mS
Temp. ref.
25°C
1413 µS
12.88 mS
111.8 mS
4.0 mS
Calibration TC:
KCl liquid TC