EC-METER EC1200 AND PH-METER PH1200
_____________________________________________________________________________________________
15
Manual temperature
This calibration can be done when the Pt1000 sensor is not connected or malfunctioning and the user
wants to adjust the manual temperature compensation value.
[MODE
] push the key from the main display in order to reach the D2.0 display.
[SHIFT]
push the key to read the manual temperature value with the flashing cursor on the last digit
XXX.X °CM
or (°FM)
[INCR]
push the key to modify the last digit if necessary
[SHIFT]
push the key to position the cursor on the other digit to be modified with same procedure.
[ENTER]
push the key for 3 seconds to confirm the adjusted value.
At this point the procedure is same as the above steps.
The message
Update
indicates the new calibration values have been memorized.
If the new value is out of the accepted limits, the display will show the message
min/Max 0/100 °C
or
32/212 °F
for few seconds.