background image

Kestrel

®

 4200 Pocket Air Flow

®

 Tracker

13

the altitude readings tend to fluctuate by a few feet.) After obtaining a current altitude from the ALTITUDE 

screen, move to the BARO screen and enter this value as your reference altitude by following the same 

procedure. Both readings will now be accurate.

Starting with a known altitude for your location

 

You can obtain your altitude from a topographical map or local landmark.  Google Earth  

 

is an excellent free program that provides the exact altitude for any given address:  

 

www.earth.google.com/. Set this value as your reference altitude on the BARO screen  

 

to determine your barometric pressure: Press the   button to enter the reference  

 

setting mode. Press the   button to increase the reference altitude or the   button to 

decrease the reference altitude. You will notice that the barometric pressure will change with changes in 

the reference altitude. Press the   button to exit the adjustment mode. Again, allow the Kestrel Meter to 

stabilize, then enter the value from the BARO screen as your reference pressure on the ALTITUDE screen by 

following the same procedure. Both readings are now accurate.
When reviewing stored data, remember that changes in pressure AND changes in location/altitude will 

affect the stored values. When tracking pressure changes relative to weather, set the reference altitude on 

the BARO screen and keep the Kestrel Meter in one location. Your graph history will now show trends in 

barometric pressure. Your altitude as shown on the ALTITUDE screen will change as the weather changes, 

but you can ignore this screen for this purpose.
If you are planning a day hike would like to track your altitude, you’ll need to enter the correct reference 

pressure on the ALTITUDE screen as described above in “starting with the known barometric pressure.”   

You can now track the altitude changes as you hike. In this instance, you should ignore the values on the 

BARO screen, since the pressure changes will be due to changes in elevation far more than to changes in 

the weather.

In general, changes in barometric pressure associated with weather changes are small over the course of 

one day, but they will affect the accuracy of the altimeter over time. This is why aircraft reset their altimeters 

at every airfield by entering the field’s “altimeter setting” or reference pressure. Accordingly, if accurate 

altitude readings are your primary interest, you should reset the reference pressure on your Kestrel Meter 

regularly. If you encounter an elevation landmark, you can adjust the reference pressure until the altitude 

matches the landmark elevation. This will correct the altitude for any pressure changes due to the weather. 

(Or, you can obtain an updated reference pressure from the sources described above.)
Some final notes — If you wish to know the actual or station pressure for your location (such as for engine 

tuning), simply set the reference altitude on the BARO screen to “0”.  In this case, the Kestrel Meter will 

not make any adjustment and will display the measured value.  (Engine tuning and ballistics software 

sometimes refer to atmospheric or station pressure as “absolute pressure.”  These applications are 

concerned with the actual air density, as opposed to pressure gradients relating to weather, so barometric 

pressure is less useful.)
Also, the above discussion applies to ALL pressure altimeters, including one you may have in a watch or 

other device, but not to GPS altimeters, which use satellite triangulation to determine altitude. Note that 

with present GPS technology, pressure altimeters remain more accurate for measuring altitude change.  

This is why airplanes still rely on pressure altimeters, not GPS.
Finally, the DENSITY ALTITUDE screen is calculated from the absolute values of station pressure,  

relative humidity and temperature, and is not affected by the reference values entered in the BARO and 

ALTITUDE screens. 

Shop for Kestrel products online at:

1.877.766.5412

www.

KestrelStore

.com

Summary of Contents for Kestrel 4200

Page 1: ...acker 4200 Pocket Air FlowTracker with Backlight In a hurry Be sure to read pages 5 7 to ensure that your Kestrel is setup properly www kestrelweather com 800 784 4221 Revised 07 28 10 Shop for Kestre...

Page 2: ...to enter and exit Main Setup Menu IMPELLER Zytec V bearings on a user replaceable impeller IMPELLER COVER Swivel cover protects impeller when not in use DATA UPLOAD OPTICAL COUPLER Software and seria...

Page 3: ...rd through the loop on the thin end Using tweezers can help Battery Installation Use only AAA batteries Install batteries as indicated on the battery door After installing the batteries the Kestrel 42...

Page 4: ...hat the cursor will remain at the same date and time If new data is stored while viewing chart data the entire chart will shift left with the new data point charted on the right The cursor will not sh...

Page 5: ...minute Press the button within one minute to deactivate the light manually Airflow The Kestrel Air Flow Tracker will display the volume airflow based on the instantaneous air velocity and cross secti...

Page 6: ...that point the RH value is stable and can be relied upon to be within the accuracy specifications Barometric Pressure and Altitude Adjustment The Kestrel Meter measures station pressure the actual air...

Page 7: ...ou should ignore the values on the BARO screen since the pressure changes will be due to changes in elevation far more than to changes in the weather In general changes in barometric pressure associat...

Page 8: ...Max Avg data is cleared Chart data will remain intact Press or to clear the MMA Auto Store On Off When On data is automatically stored at preset Store Rate When Off data is only stored when manually c...

Page 9: ...etup Menu User Screens The three User Screens can be reconfigured to display the most appropriate information for your application For example if you need to monitor the wind speed humidity and barome...

Page 10: ...ion for instructions Then adjust the calibrating setting on the Baro Cal screen to the known mean sea level barometric pressure If recalibration is desired the unit may also be returned to NK for serv...

Page 11: ...xamine the effect on the environment when employees enter and exit the building HVAC R System Balancing Auto Store Off Store Rate Overwrite Off Man Store On These settings will require you to press th...

Page 12: ...into your location Falling pressure indicates the arrival of a low pressure system and expected precipitation or storm conditions Steady or rising pressure indicates clear weather A correct altitude...

Page 13: ...physiology guides use wet bulb temperature rather than heat index as a measure of the safety of exercise in hot and humid conditions Wind Chill The cooling effect of combining wind and temperature Th...

Page 14: ...ic device and it will display a wind speed Or wave a magnet like a refrigerator magnet back and forth by the Kestrel Meter If the Kestrel Meter is working properly you will register a wind speed readi...

Page 15: ...ue to improper use or neglect including corrosion damage caused by severe or excessive impact crushing or me chanical harm modifications or attempted repairs by someone other than an authorized NK Kes...

Page 16: ...es Incoming and outgoing data is supplied with the certificate of calibration We also offer full factory service on every product we manufacture for as long as we make the product and as long after as...

Page 17: ...tion Manual for Kestrel 4200 version 4 29 ALL NIELSEN KELLERMAN 21 Creek Circle Boothwyn PA 19061 Phone 610 447 1555 Fax 610 447 1577 Web www kestrelweather com Email kestrel nkhome com Please registe...

Reviews: