background image

1. LCD SCREEN STANDARD VIEW

Test Mode Icons:

2. Battery icon: Battery level or charging status. Flashes when

gauge needs to be recharged.

3. OK/OV Indicator:

4. Force Icons: Indicates force direction.

5. Current measured value

6. Analog bar: Indicates current position within full scale. When

the bar enters the area enclosed by the dotted line, this

signifies the full scale capacity is exceeded by an overload

condition.

7. Storage icon: Indicates data is being saved.

8. System time

9. Units Indicator: Selected engineering unit.

10. Load Cell Capacity Icon:

2

S-Beam for Compression or Tension

N

kgf

lbf

kN

S-1

1kN

Capacity

1000

100

220

-

Resolution

0.1

0.01

0.05

-

S-2

2kN

Capacity

2000

200

450

-

Resolution

0.5

0.05

0.1

-

S-5

5kN

Capacity

5000

500

1100

-

Resolution

1

0.1

0.1

-

S-10

10kN

Capacity

10000

1000

2250

10

Resolution

1

0.1

0.5

0.01

S-20

20kN

Capacity

-

2000

4500

20

Resolution

-

0.5

1

0.005

Ring Type for Compression

N

kgf

lbf

kN

klbf

tf

R-10

10kN

Capacity

10000

1000

2250

10

-

-

Resolution

0.1

0.1

0.5

0.01

-

-

R-20

20kN

Capacity

-

2000

4500

20

-

-

Resolution

-

0.5

1

0.005

-

-

R-50

50kN

Capacity

-

5000

-

50

11

5

Resolution

-

1

-

0.01

0.001

0.001

R-100

100kN

Capacity

-

-

-

100

22

10

Resolution

-

-

-

0.01

0.005

0.001

R-200

200kN

Capacity

-

-

-

200

44

20

Resolution

-

-

-

0.05

0.01

0.005

1

2

3

4

9

8

7

6

5

10

Under Lower Limit

Between Low Limit & Upper Limit

Over Upper Limit

Tension

Compression

Track: Real Time, live measuring mode

Peak: Reading will not change until a higher value is mea-
sured
AutoPeak: When Peak Time is up, resets peak value automat-
ically
First Peak: Captures First Peak after drop ratio decrease has 
been detected. Drop Ratio set in menus.

If no load cell is connected, this symbol appears & blinks

2. LOAD CELL CAPACITY & RESOLUTION TABLES

Digital View

Graphic View

1

3

5

9

8

2

7

10

6

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Shimpo FG-7000L-S-5

Page 1: ...rs should wear protection such as a mask and gloves in case pieces or components break away from the unit under test Whether the unit is ON or OFF DO NOT exceed the capacity of the sensor NEVER exceed...

Page 2: ...0 5 0 01 S 20 20kN Capacity 2000 4500 20 Resolution 0 5 1 0 005 Ring Type for Compression N kgf lbf kN klbf tf R 10 10kN Capacity 10000 1000 2250 10 Resolution 0 1 0 1 0 5 0 01 R 20 20kN Capacity 2000...

Page 3: ...to cancel or touch MENU to confirm and exit Figure 4 2c Figure 4 2c 4 2 2 Group When several test samples need to be measured the sam ples can be coded into groups The range is 01 99 When set to 00 b...

Page 4: ...ENU to confirm and exit Figure 4 3b Single At the home screen pressing the LOG stores the cur rent displayed value If the default settings key is for storage See 4 5 8 key setting Series Continuous da...

Page 5: ...rs may be set per Fig ure 4 5a 4 5b Figure 4 5a Figure 4 5b 4 5 1 Display Mode Two display modes may be selected Digital and Graphic Fig ure 4 5c Figure 4 5c 4 5 2 Display Direction Select the mode of...

Page 6: ...can restore it to the factory default settings Carefully use this function Figure 4 5l Figure 4 5l 4 6 Language Select between English German and Chinese Figure 4 6a Figure 4 6a 4 7 Calibration Users...

Page 7: ...1a The USB Port can be connected to a charger with the USB cable for charging the internal Ni MH battery and can be connected to a computer for uploading the measured values Connect the gauge and the...

Page 8: ...ee Figure 6 1a 7 3 Dimensions S Beam Ring Type Capacity Dimensions W H B M 1kN 51 76 2 19 M12X1 75 2kN 5kN 10kN 25 4 20kN 76 2 108 M18X1 5 Capacity Dimensions mm D d L H SR DI M 10kN 32 8 14 16 16 24...