■ Measuring procedure
Install the stand body on a level solid table.
During measurement, operate the lever at a constant speed as smoothly as possible. If the lever operation speed is rapidly
changed (to a high or low speed), accurate measurement cannot be performed.
During measurement and height adjustment, use thorough caution not to apply impact to the force gauge measuring axis.
The measuring stand can be horizontally installed.
To apply large load, fasten the stand body.
For details on the force gauge, refer to the force gauge operation manual for each model.
■ Stopper adjustment procedure
To determine the force gauge lower end position during measurement,
adjust the stopper position according to the procedure below.
Loosen the lock nut at the bottom of the slide assembly.
1) As the adjusting bolt is screwed into the slide assembly, the stopper
position is raised.
2) After positioning the stopper, tighten the lock nut to complete the
stopper adjustment procedure.
■ Force gauge mounting procedure
1) Remove the mounting bolts by using the hexagon wrench (accessory), and
remove the force gauge mounting plate.
2) Mount the force gauge to the force gauge mounting plate as shown below.
3) Mount the force gauge to the stand body by using the removed bolts.
FGJN/FGP/FGPX series
Cross-recessed
pan head screw,
M4 x 8 (4 pcs)
(accessory)
FGX
(old model)
series
Cross-recessed
pan head screw,
M 5 x 6 ( 3 p c s )
(accessory)
Force gauge
Mounting plate
Mounting plate
Mounting plate
Slide assembly
Adjusting range
Lock nut
Adjusting bolt
Force gauge