OPERATING INSTRUCTIONS
SETTING TEMPERATURE
1. Press and hold the DOWN arrow for more than 1.5 seconds or press the
OR
FUNCTION button to display “SP1” the set temperature.
The 'set temperature' indicator will illuminate.
2. Use UP and DOWN arrow keys to select required temperature.
3. After setting temperature the display flashes between “SP1” and set
temperature and will automatically revert to show actual liquid temperature.
Heater indicator will illuminate.
4. The over or under temperature alarm is automatically set 4°C above and
4°C below set temperature.
If the actual temperature rises or falls beyond
this value the audible alarm and indicator are automatically activated.
5. The Clifton water bath is now set and will heat and control the liquid to set
temperature.
The Clifton Range - 3 Year Warranty
10
High Quality Products at Affordable Prices
SETTING TIME
1. Press FUNCTION button until "t" appears on the display. It will
then alternate between showing "t" and time - displayed as h:mm.
2. Press either UP or DOWN arrow to select desired number of hours.minutes.
- Minimum time setting is 0 hours and 01 minutes - displayed as 00.01
- Maximum time setting is 9 hours and 59 minutes - displayed as 9.59
3. Once the desired time is entered press the function button to confirm.
Display reverts to actual liquid temperature.
TIMER OPERATION
1.To start the timer press the RUN button for more than 1.5 seconds. The
Timer automatically starts once bath temperature is at set temperature.
2.Heating indicator will illuminate continously while liquid temperature is
being raised to set point.
3.Heating indicator will flash when set temperature has been reached and
is being maintained.
4.The timer indicator starts to flash showing the timer is running.
5. To view time remaining, press UP ARROW button until 't'
appears and time remaining is displayed as h:mm.
6. An audible beeping and "End" message indicates timed period has finished.
Press RUN button to deactivate buzzer and clear "End" message.