Service manual
Reducing Vaporization Mercury Analyzer Mercury RA-4500
18
(5) Remove the pump unit from the cabinet.
(6) When the removal is difficult, disconnect
the air pump and reagent pump
connectors.
(7) Disconnect the “COLOR” connector.
(8) Remove the screws fixing the dispensing
case.
(9) Remove the connectors for the
dispensing unit.
(10) Remove the dispensing case.